15
Uncover the sensor and wait for the shortwave radiation reading to
settle
before
entering the reference value. Click OK.
The multiplier and offset values will automatically calculate and
update in the appropriate field. Be sure to click ‘Save’ to retain the
new multiplier and offset.
Clicking
‘Data Logging’
will allow the user to log interval
measurements in a csv file while the software is open and
communicating with the sensor.
Click ‘Setup’ and the Setup Logging window appears. Click the
‘Browse’ button to create or select a csv file.
Select the desired sampling interval. Note that 1 second is the
minimum interval allowed. Click ‘Start’.