33
ELECTROMAGNETIC IMMUNITY
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Electrostatic Discharge (ESD)
EN 61000-4-2
± 2, 4, 6 kV
contact discharge
± 2, 4, 6 kV
contact discharge
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
± 2, 4, 8 kV
air discharge
± 2, 4, 8 kV
air discharge
Electrical Fast
Transient/Burst
EN 61000-4-4
±2 kV AC Mains
±1 kV I/O Lines
±2 kV AC Mains
±1 kV I/O Lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
EN 61000-4-5
±1 kV
Line to Line
±2 kV
Line to Ground
±1 kV
Line to Line
±2 kV
Line to Ground
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
If the user of the Winx Sleep Therapy System requires
continued operation during power mains interruptions, it
is recommended that the Winx Sleep Therapy System be
powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
EN-61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
PN 12149-B.indd 33
4/23/13 11:53 AM
Summary of Contents for Winx
Page 1: ...Winx Sleep Therapy System User Guide for Patients PN 12149 B indd 1 4 23 13 11 53 AM...
Page 2: ...PN 12149 B indd 2 4 23 13 11 53 AM...
Page 4: ...PN 12149 B indd 4 4 23 13 11 53 AM...
Page 6: ...PN 12149 B indd 6 4 23 13 11 53 AM...
Page 14: ...8 PN 12149 B indd 8 4 23 13 11 53 AM...
Page 22: ...16 PN 12149 B indd 16 4 23 13 11 53 AM...
Page 30: ...24 PN 12149 B indd 24 4 23 13 11 53 AM...
Page 36: ...30 PN 12149 B indd 30 4 23 13 11 53 AM...
Page 46: ...40 PN 12149 B indd 40 4 23 13 11 53 AM...
Page 57: ...51 Notes PN 12149 B indd 51 4 23 13 11 53 AM...
Page 58: ...52 Notes PN 12149 B indd 52 4 23 13 11 53 AM...