27
Problem
Possible Cause
Solution
Cannot feel
vacuum
in mouth/
Cannot
maintain
target
vacuum ( )
Mouthpiece not
sitting correctly
in mouth or
not providing a
proper seal
• Make sure lip seal is positioned between lips and teeth.
• Make sure there are no tears or other damage to the lip seal.
• Make sure the mouthpiece is oriented correctly within your
mouth. The arched pad should be facing up toward the
roof of your mouth, and the pad should rest in the back of
your mouth.
• Make sure mouthpiece is attached to the tubing.
Console/tubing
leak
• Make sure all connections are properly seated.
• Make sure the lid on the reservoir is properly sealed and that
the reservoir is completely inserted into the console.
• Check tubing for cracks/leaks. If you find a crack or leak,
replace the tubing.
System blocked
• Check that the inside of the reservoir lid has been
thoroughly rinsed and dried. Allow lid to air dry for at least
five minutes before replacing.
• Check tubing for kinks or knots, and straighten if necessary.
PN 12149-B.indd 27
4/23/13 11:53 AM
Summary of Contents for Winx
Page 1: ...Winx Sleep Therapy System User Guide for Patients PN 12149 B indd 1 4 23 13 11 53 AM...
Page 2: ...PN 12149 B indd 2 4 23 13 11 53 AM...
Page 4: ...PN 12149 B indd 4 4 23 13 11 53 AM...
Page 6: ...PN 12149 B indd 6 4 23 13 11 53 AM...
Page 14: ...8 PN 12149 B indd 8 4 23 13 11 53 AM...
Page 22: ...16 PN 12149 B indd 16 4 23 13 11 53 AM...
Page 30: ...24 PN 12149 B indd 24 4 23 13 11 53 AM...
Page 36: ...30 PN 12149 B indd 30 4 23 13 11 53 AM...
Page 46: ...40 PN 12149 B indd 40 4 23 13 11 53 AM...
Page 57: ...51 Notes PN 12149 B indd 51 4 23 13 11 53 AM...
Page 58: ...52 Notes PN 12149 B indd 52 4 23 13 11 53 AM...