PG5
Rev. A1, 3/10
16
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]
Full-Scale Range Adjust (RANGE):
Allows the user to adjust the reading at full-scale pressure. The reading can be
adjusted by +/-10% full-scale.
NOTE:
The pressure reading must be within 5% of the full-scale value in order
to make Range adjustments. For example, a 1000 psi gauge would need
to be reading between 950 psi and 1050 psi in order to adjust the Range
feature. If the reading is not within 5% of full scale, NOAdJU (No
Adjustment) will be displayed when trying to adjust the Range.
Adjusting the Full-Scale Range:
Step 1: Ensure that the pressure reading is within 5% of full-scale.
Step 2: Press the On/Off button to enter the main setup menu.
Step 3: Press (Z) to cycle through the options until AdVSET is displayed.
Step 4: Press (S) to access the Advanced Settings menu.
Step 5: Press (Z) to cycle through the options until RANGE is displayed..
Step 6: Press (S) to enter the Range adjust mode.
Step 7: Press (Z) to increase the reading, or press (S) to decrease the reading.
Step 8: Press the On/Off button to accept the adjusted reading and return to
the main setup menu.