40417 - Offset (Factory Calibrated)
Sets the Zero Reference of the sensor, the point from which the calculated distance is measured.
40418 - Pre filter
Defines the physical range (0 - 10,364 mm) of the start up (pre-filter) window. Four sample readings must be
found within the Pre filter window for the MPXI-F sensor to successfully start up.
This register is only to be used for diagnostics under factory direction.
40419 - Noise limit
Sets the limit for number of signals (0-255) outside the Pre filter range for the MPXI-F at start up. If the Noise
Limit is reached before four readings register within the Pre filter window, the MPXI-F will not start up.
This register is only to be used for diagnostics under factory direction.
40420 - Temperature Select
Selects the temperature sensor reading to be displayed in Input Register 30302.
MPXI-F8 sensors can accomodate up to seven digital temperature sensors in the stem.
0 = Average of sensors A - G
1 = Digital Temperature Sensor A
5 = Digital Temperature Sensor E
2 = Digital Temperature Sensor B
6 = Digital Temperature Sensor F
3 = Digital Temperature Sensor C
7 = Digital Temperature Sensor G
4 = Digital Temperature Sensor D
8 = N/A