MultiSure GK Link Blood Glucose and Ketone Monitoring System
46
• Once CTL mode is enabled, CTL mark is displayed together with “ctl” message
written in bigger characters on the main screen.
Step 3.
Gently shake the control solution vial before testing. Discard a drop before use.
Squeeze a drop of control solution (Level 1, Level 2 or Level 3) onto a clean, hard,
dry nonabsorbent surface. Do not apply control solution to the test strip directly from
the bottle. Replace the bottle cap on the control solution bottle immediately after use.
Step 4.
Hold the meter and touch the control solution to the edge of
the sampling end of the test strip. The control solution will be
automatically pulled into the reaction area of the test strip. A beeper
will sound (if enabled) as the countdown timer starts on the screen.
NOTE:
• The test will not start if you apply your control solution not directly to the edge of the
sampling end of the test strip. The test starts when the meter detects the control
solution. During the test the meter counts down from 5 to 1.
• Tightly close the control solution vial.
• Do not touch the test strip once the meter has started the countdown.