background image

HumaSens2.0

plus

 

Starter Pack

 

  

 User Manual

Summary of Contents for HumaSens2.0plus

Page 1: ...HumaSens2 0plus Starter Pack User Manual...

Page 2: ..._______________ 5 Test Strip and Code Strip_________________________________________________________ 6 Chapter 2 Setting the Meter 2 1 Installing Replacing the Batteries_______________________________...

Page 3: ..._________ 19 Chapter 7 Error Message and Trouble Shooting__________________________________________ 20 21 Chapter 8 Specifications________________________________________________________________ 22 Ch...

Page 4: ...blood glucose total cholesterol and uric acid levels in human capillary whole blood from the finger prick It is used to monitor the level of glucose total cholesterol and uric acid The HumaSens2 0plu...

Page 5: ..._____ 1 Optional not included in the standard kit package please contact your local distributor for ordering REF Glucose Test Strips Blue HumaSens2 0 HumaSens2 0plus HumaSens HumaSensPlus 50 2 x 25 pc...

Page 6: ...ure limitation Manufacturer Use within 3 months after initial opening Storage and transpor tation temperature limitation Do not reuse Attention Storage and transpor tation relative humidity limitation...

Page 7: ...results Right S Button Confirm and set up Left M Button Enter memory mode Press long to turn off Ejection Button Remove test strip Test Port Insert test strip Serial Number Sticker Battery 3V lithium...

Page 8: ...nd test result display 1 4 Alert Tones Normal Alert a short beep Warning Alert 3 short beeps Turning On Off a long beep 1 Result Area 2 Low Battery 3 Temperature Icon 4 Average GLU 5 Unit Icon 6 Time...

Page 9: ...coding step when you use a new vial of strip Electronic Contact Bars Insert this end into meter test port Contact Bar Insert this end into the test port of the meter Code Test Mode Sample Inlet The w...

Page 10: ...s according to your local environmental regulations 2 2 Set Date Time and Units Press S button for over three seconds You will hear one beep for entering the setting mode then follow the instruction b...

Page 11: ...new vial of strips With the Auto Strip Identification function once you code your meter you do not have to switch the test mode every time you perform a test Make sure the meter is off Then insert the...

Page 12: ...need to perform a test HumaSens2 0plus Meter HumaSens2 0plus Test Strip Lancing Device Lancets not provided Tissue or cotton ball with 75 ethanol or disinfection wipes for cleaning the sampling area b...

Page 13: ...appear remove the unused strip and repeat step 2 4 The pa tient s finger must be warm dry at sur face clean including being free from sub stances such as hand cream Use a single use lancet not provide...

Page 14: ...t fill up of blood sample could lead to inaccurate or failed test result After count down of the respective tests time need seconds the result will display on the screen Then eject the used strip into...

Page 15: ...es or the meter will turn off automatically The meter will not turn on if you insert the wrong end or wrong side of strip Do not use expired strip Apply the blood sample to the correct position on the...

Page 16: ...not provided into biohazard waste container 2 Record your test result 3 Store your instrument strips and other items in a clean cool and dry environment Please check the specifications in Chapter 8 T...

Page 17: ...agnetic and radioactive interferences Warning Do not disassemble the meter for any reason Please follow local regulations to discard used test strips lancets and batteries Used test strips lancets and...

Page 18: ...for Checking Memories 1 No test strip in the meter 2 Press M button for 3 seconds till a short beep to enter the memory mode 3 After a full display on screen press M button to switch between GLU CHOL...

Page 19: ...16 GLU Average CHOL or UA Individual Record CHOL or UA Control Record GLU Individual Record GLU Control Record...

Page 20: ...hould be checked If you suspect that the meter or test strips are not working properly When the test result is or seems unusual inaccurate or inconsistent Any time you droped or bumped the meter Contr...

Page 21: ...inserts for HumaSens2 0plus strip and HumaSens2 0plus control solution before use Do not reuse the test strip Newly opened vials of control solution and test strips must be marked with the opening dat...

Page 22: ...or expose them directly to the sunlight Do not freeze or refrigerate the meter and strips Keep the meter clean by wiping the exterior surfaces with tissues or lint free cloth Keep your and the patient...

Page 23: ...immediately Problem with code strip Code strip inserted in an improper way Repeat the coding procedure and ensure the code number is the same in label code strip and screen If the problem persists ple...

Page 24: ...n label code strip and screen If the problem persists please contact the distributor for service Not enough sample volume for the test Take a new strip and make sure you have enough sample volume for...

Page 25: ...nds UA 15 seconds Measuring Range GLU 20 600 mg dl 1 1 33 3 mmol l CHOL 100 400 mg dl 2 59 10 35 mmol l UA 3 20 mg dl 0 18 1 19 mmol l Sample Volume GLU 0 7 l CHOL 10 l UA 1 l Hematocrit Range 25 60 G...

Page 26: ...rnal use only HumaSens2 0 HumaSens2 0plus test strips are designed for single use only Do not reuse Do not perform a test while a mobile phone is used in the vicinity to prevent potential electronic s...

Page 27: ...Lipemic samples Cholesterol levels up to 326 mg dl 8 43 mmol l do not affect the results Grossly lipemic patient samples have not been tested and are not recommended for testing with the HumaSens2 0 H...

Page 28: ...N Gesellschaft f r Biochemica und Diagnostica mbH Max Planck Ring 21 65205 Wiesbaden Germany Tel 49 6122 9988 0 Fax 49 6122 9988 100 e mail human human de www human de 2022 HUMAN 175611 2022 05 2022 H...