The XC-3D55 3-D Flow Measurement console is factory-configured for 120VAC/60Hz and 
240VAC/50Hz electrical power. The Electrical Schematic for the console is presented on page 5. 

A front panel-mounted circuit breaker rocker switch activates the Main Power and power supply.  The 
power supply energizes the velocity pressure and pitch differential pressure transducers, the 
inclinometer (yaw measurement device) and the temperature display.  

A 4-pin amphenol-style circular connector is mounted on the side of the console is used to provide 
power to and signal from the remote inclinometer sensor.  The standard size, female, Type K Stack 
Temperature thermocouple jack is also mounted on the side of the console. 

Output signals are included for the Stack Thermocouple, Velocity Pressure, Yaw Angle and Pitch 
Angle.  All signals are direct current, voltage-based. The specific outputs are listed on the electrical 
schematic and in the specifications listed in Table 2. 

Process meters are utilized to convert the direct current voltage (VDC) output pulses from the various 
sensors into appropriate engineering units, such as inches/millimeters of water or degrees of rotation, 
etc. The process meters display the P1-P2 velocity pressure and the P3-P4 pitch angle pressure. The 
standard units of measurement are English units in “ H

2

O and 

°

 Fahrenheit. Metric units will be 

displayed in mm H

2

O and 

° 

Celsius are available.  The Yaw Angle is displayed via a high resolution 

indicator where 1 pulse equals 0.1

°

.

The process meters, transducers and counter have been setup and calibrated at Apex
Instruments utilizing appropriate standards. Read this Operators Manual before making any 
adjustments to the process meters, transducers or inclinometer. 

Main Power Switch

(M-CBR15A-M)

Velocity Pressure Display

(M-CUB5PB00)

Probe Temperature 

Thermocouple Connection

(TC-PJR)

Inclinometer Amphenol Connection

(AM-T3111)

IEC Power Input Receptacle

Pitch Pressure Display

(M-CUB5PB00)

Yaw Angle Display

(TOT4-36X72BL)

Probe Temperature Display

(TC-CUB5TCB0)

Figure 4  Electrical System Components

Electrical System Components

Revision:01                                                   Page 7 of 23                                                     9/26/2017

Summary of Contents for XC-3D55

Page 1: ...05 INTERPRET GEOMETRIC TOLERANCING PER MATERIAL FINISH CREATED CHECKED ENG APPR MFG APPR Q A COMMENTS DATE NAME TITLE SIZE DWG NO REV WEIGHT SCALE UNLESS OTHERWISE SPECIFIED SHEET 1 OF 1 DRAWING NOT T...

Page 2: ...rations do not adversely affect our warranty Important Notice This instrument provides measurement readings to its user and serves as a tool by which valuable data can be gathered The information prov...

Page 3: ...ak Check Page 15 16 System Calibration Checks Page 17 Back Purging System Page 18 21 Red Lion Cub 5 Programming Guide Page 22 Methods Comments Figures Page 4 Figure 1 Complete XC 3D55 System Page 5 Fi...

Page 4: ...easurements The wall effects adjustment factor is used with the Method 2 2G or 2F calculated velocity to derive a wall effects adjusted velocity Volumetric flow rate is calculated by multiplying the a...

Page 5: ...3D55 3 D Flow Measurement Console 1 Pitot Type 2D 3D or Type S 1 U PTX 3 D Pitot Sheath Extension 2 in Dia Sheath 1 4 in TFE Lines 1 PF 346XT32S Universal Probe Extension Flange Fits 3 4 6 in Ports 1...

Page 6: ...p Generate Pressure and Vacuum NIST 0 to 5 H2O M5C 22 Thermocouple Simulator Source 12 1057 Digital Display Reference Inclinometer Speed Square Probe Leak Check Plugs QC F4 SS Female Quick Connect wit...

Page 7: ...al Accurate within 2 range 0 5 0 0 5 w c minor divisions 02 M CUB5PB500 Process Meter 0 10VDC supply TC CUB5TCB0 Display tc pnl led 120 240V TOT4 36X72BL Totalizer quadrature 36x72 din with back light...

Page 8: ...Process meters are utilized to convert the direct current voltage VDC output pulses from the various sensors into appropriate engineering units such as inches millimeters of water or degrees of rotat...

Page 9: ...sure Pressure 1 2 3 M PSRD35A Inclinometer Amphenol Probe Temperature Display Velocity Pressure Display Pitch Angle Display Pitch Pressure Display M DP B5 B5 Main Power Figure 5 XC 3D55 Electrical Sch...

Page 10: ...ng for accurate pressure measurements Two pressure transducers measure the velocity head P1 P2 pressure and the pitch P4 P5 and are connected to LED process meters scaled and calibrated to the appropr...

Page 11: ...itot is faced into the gas stream flow and the pressure difference between the two ports equates to the velocity pressure This differential velocity pressure is measured with liquid manometers or elec...

Page 12: ...into the probe extension and over the pitot probe Check P1 Position 5 Put anti seize on the four screws and insert into the collar 6 Securely tighten the four screws 7 8 9 With the assembled pitot tu...

Page 13: ...5 Connect the Umbilical Pod Assembly UINC 3D to the XC 3D55 Console by connecting the male quick connects to the female quick connects P1 to P1 P2 to P2 P3 to P3 P4 to P4 P5 to P5 16 Insert the male t...

Page 14: ...ing using the three point calibration procedure described in section 10 3 3 of US EPA Method 2F The device shall be recalibrated according to the procedure in section 10 3 3 no later than 90 days afte...

Page 15: ...t until at least 7 6 cm H2O 3 in H2O pressure or a pressure corresponding to approximately 75 percent of the pressure measuring device s measurement scale whichever is less restrictive registers on th...

Page 16: ...from the console 2 Press Reset to Zero 3 Connect the pressure measuring device with pressure generator to P1 Port 4 Vent all ports Let devices settle Record values for ZERO 50 of full range Value mus...

Page 17: ...acceptable limit consult the Calibration Procedures section 4 Record the reading at each calibration check point Calibration Checks of Angle Measuring Device 1 Securely attach the calibrated Referenc...

Page 18: ...edone Note that the selected traverse point should be one that demonstrates a readable p value If back purging at regular intervals is part of a routine procedure then comparative p measurements shall...

Page 19: ...5 Thermocouple Meter Offset Guide TC CUB5TCB0 Press and hold SEL for 2 seconds Display will alternate from Pro to NO screen Press RST once Your screen should switch to1 InP screen Press SEL 5 times Di...

Page 20: ...f 0 000 follow next 2 steps Press RST until each value equals 0 Hold SEL for 2 seconds to move to Inp 2 Press RST once to unlock the value Press RST once to unlock the value for change Press SEL 3 tim...

Page 21: ...s and hold SEL for 2 seconds Press RST 3 times to 3 DSP Press SEL 5 times to Pro Code Press RST to scroll through numbers and select 1st number Press SEL to lock the 1st number The display will automa...

Page 22: ...change 2nd number 022 222 Press SEL to change 3rd number Press and hold SEL for 2 seconds and follow procedure for programming Pro NO Press RST repeatedly until reaching desired 1st number Press RST...

Page 23: ...igure 10 Method 2 Determine the atmospheric pressure at the sampling elevation during each test run following the procedure described in section 2 5 of Method 2 Method 3 Determine the stack gas dry mo...

Page 24: ...os End View of Probe Direction of Flow Figure 11 Probe Yaw Orientation Positive Negative Figure 10 Method 1 Traverse Point Determination Marks Mark Traverse Points on Probe according to Method 1 Marks...

Reviews: