Calibration Checks of Temperature Measuring Device
1) Connect the NIST traceable Reference Temperature generating device to the Thermocouple Jack
on the console.
2) Starting at ZERO
°
F, check the temperature display on the console at 0
°F
, 100
°F
, 200
°F
, 300
°F
,
400
°F
and 500
°
F. The console display temperature reading must be +/-5
°
F of the reference reading.
Should any reading FAIL the acceptable limit, consult the Calibration Procedures section.
4) Record the reading at each calibration check point.
Calibration Checks of Angle Measuring Device
1) Securely attach the calibrated Reference Inclinometer to the P1 Velocity Pressure orifice flat on the
Pitot Probe Head. Insure the UINC-3D-## assembly is securely attached to the Probe Extension and
Pitot..
2) Rotate the Reference Inclinometer to a desired position and record the angle.
3) Read the Yaw Angle indicated on the console display and press reset.
2
1
3
2
1
3
2
1
3
2
1
3
Revision:01 Page 16 of 23 9/26/2017