OXYGEN ANALYZER
USING THE O2 ANALYZER WITH THE DSX
First plug the O2 Analyzer cable into the DSX. To access the O2 Analyzer menu, hold the Left and Right buttons
while on the Surface Main screen in CC, OC, or SM Mode. Alternately, O2 Analyzer is an option in the Set Gases
Menu of those modes, see p. 32. The first step to use the analyzer is to calibrate the sensor.
O2 ANALYZER
/
CALIBRATE
ANALYZE
/
WARNING
USE ONLY WITH
O2-000000
CALIBRATION
Changing enviornmental conditions and the diminishing output of an O
2
sensor can affect the gas analysis. For
this reason it is good practice to calibrate periodically, minimum of every day of use. Set the FO
2
(Fraction of Oxy-
gen) of the known gas.
1. Hold the O2 Analyzer dome cap firmly against the known (control) gas tank valve or expose in air. Slowly open
the tank valve only part way if using a cylinder gas to calibrate with. The flow rate should not be excessive (≤ 10
liters per minute).
2. Expose the sensor until the DSX confirms the calibration.
3. If any error codes are displayed, take corrective action before attempting to calibrate again.
4. After successful calibration you may use the Analyze feature.
!
DANGER: Moisture in the gas can create false readings.
!
DANGER: DO NOT proceed until a good calibration is achieved.
NOTES: DO NOT calibrate with excessive flow rates (greater than 10 liters per minute).
NOTES: Dropping or mishandling the O2 Analyzer can damage the O
2
Sensor or electronics. It is also
important to keep the O2 Analyzer dry.
CALIBRATE
/
98
/
SET CALIBRATION
GAS FO2
CALIBRATE
/
SURFACE PRESSURE
1012 mBar
READY FOR
CALIBRATION
CALIBRATE
READING
O2 CELL
VOLTAGE
SELECT THE CALIBRATION
GAS
START GAS FLOW & START
CALIBRATION
CONTINUE HOLDING DOME CAP
AGAINST TANK VALVE
!
DANGER: Your new DSX and O2 Analyzer cable are paired and
precisely tuned at the factory, to compensate for resistance in the
cable. ONLY use the cable that ships with your DSX. The cable
serial number (labeled on the cable) must match the serial number
that displays on your DSX (O2 Analyzer Sub Menu) screen. Contact
your Apeks Dealer for assistance if the Oxygen Analyzer cable ever
needs to be replaced.
CALIBRATE
CALIBRATED BASED ON
FO2: 98
CELL VOLTAGE
46 mV
CALIBRATE
O2 CELL
CALIBRATION
COMPLETE
CALIBRATION REPORT
SUCCESSFUL CALIBRATION
Summary of Contents for NS158000
Page 8: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 8 Apeks 2022 GETTING STARTED...
Page 14: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 14 Apeks 2022 DIVE FEATURES...
Page 23: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 23 Apeks 2022 CC CLOSED CIRCUIT SURFACE MODE...
Page 46: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 46 Apeks 2022 CC CLOSED CIRCUIT DIVE OPERATION...
Page 59: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 59 Apeks 2022 OC OPEN CIRCUIT MODE...
Page 65: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 65 Apeks 2022 GAUGE MODE...
Page 68: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 68 Apeks 2022 SM SIDEMOUNT MODES...
Page 73: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 73 Apeks 2022 NAVIGATION...
Page 80: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 80 Apeks 2022 OXYGEN ANALYZER...
Page 87: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 87 Apeks 2022 REFERENCE...
Page 90: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 90 Apeks 2022 TECHNICAL DATA...
Page 96: ...Doc 12 7969 Doc 12 7969 r09 9 22 22 96 Apeks 2022 www apeksdiving com...