APAR AR233 User Instruction Download Page 8

8

 

Table 9. List of configuration parameters 

No. 

Name 

Description of parameter 

User 

settings

 

RTC

    (1)

 

internal real time clock (chapter 9.1) 

 

RECORDING OPTIONS

 (chapter 9.2)

 

Start

 

recording start time 

 

Stop

 

recording end time 

 

MPer 

data recording interval 

 

SDest 

selection of memory to save archive files (to record data) 

 

CONFIGURATION OF THE MEASUREMENT INPUT

 (chapter 9.3) 

inp 

type of measurement input 

 

rrtd 

total resistance of leads for 2-wire RTD sensors and 700 Ω 

 

cjte 

temperature of cold thermocouple tips 

 

dot 

position of the point/resolution 

 

rbot 

bottom of the indication range (start of the scale) for 0 mA, 4 mA, 0 V, 0 mV, and 0 Ω 

 

10 

rtop 

top of the indication range (end of the scale) for 20 mA, 10 V, 60 mV, and 700 Ω 

 

11 

calO 

calibration of the zero 

 

12 

calG 

calibration of the inclination 

 

ACCESS OPTIONS

 (chapter 9.4) 

13 

bloc 

block of the “

COPY

” button

 

 

14 

Pass 

card authorization password

 

 

15 

Prot 

SD card authorization demand 

 

16 

ID       

(1)

 

identification number, ID (recorded in the csv archive file) 

 

Notes: (1)

 – parameter accessible only from the level of ARSOFT-CFG WZ1 (on-line configuration) 

 
 

9.1. TIME AND DATE 

The current time and date are used as time stamps for recording. Their review and modification are possible 

only

 

from the level of the ARSOFT-CFG WZ1 software (item 1 of chapter 9). 

Table 9.1. Parameters of the real time clock (

RTC

Parameter

 

Range of variability of the parameter 

Time (hh:mm:ss) 

00:00:00

 ÷ 

23:59:59

 

Date (dd:mm:yyyy)

 

01.06.2008

 ÷ 

31.12.2099

 

 
 

9.2. RECORDING OPTIONS 

Data is archived in a text file with a “

csv

” extension in the internal memory, on an SD/MMC card; a detailed 

description of the storage format is given in chapter 11. 

Recording to a file is performed only when the current 

time (

RTC

) is within a range defined by recording parameters 1:

Start

 

and 2:

Stop

 

(Table 9.2). 

Recording continues 

until the memory becomes full and then is stopped. The archive files must then be transferred or copied for 
further analysis so as to free up space in the memory for new recording. Selection of the memory for recording is 
determined by parameter 4:

SDest

 

 

Summary of Contents for AR233

Page 1: ...USER INSTRUCTION DATA RECORDER AR233 Version 1 0 5 2016 07 07...

Page 2: ...ASUREMENT CONNECTOR DESCRIPTIONAND THE ELECTRICAL CONNECTIONS 6 7 CONNECTING TOA COMPUTERAND INSTALLATION OF DRIVERS 6 8 INSTALLATION OF SOFTWARE 7 9 SETTING OF THE CONFIGURATION PARAMETERS 7 9 1 TIME...

Page 3: ...ity that the user changes the battery long operation time with a new battery up to 5 years depending on the recording interval the presence of an SD card and the operating temperature of the device po...

Page 4: ...850 C Ni100 RTD 3 or 2 wire 50 170 C thermocouple J 40 800 C thermocouple K 40 1200 C thermocouple S 40 1600 C thermocouple B 300 1800 C thermocouple R 40 1600 C thermocouple T 25 350 C thermocouple E...

Page 5: ...lity EMC immunity according to the PN EN 61000 6 2 2002 U standard emissivity according to the PN EN 61000 6 3 2002 U standard 1 operation time depends on the measurement period the presence of an SD...

Page 6: ...e View button in the dialog box select the DRIVERS folder press the Continue button when the warning Installing a device appears press Continue anyway 4 The virtual COM port CDC USB to UART is install...

Page 7: ...or or analog signal must be connected and the configuration must be programmed There are parameter configuration methods 1 Use the USB port and the ARSOFT CFG WZ1 software on line configuration connec...

Page 8: ...d 16 ID 1 identification number ID recorded in the csv archive file Notes 1 parameter accessible only from the level of ARSOFT CFG WZ1 on line configuration 9 1 TIMEAND DATE The current time and date...

Page 9: ...np type of input Pt100 thermoresistance sensor RTD Pt100 200 850 C Pt100 Ni100 thermoresistance sensor RTD Ni100 50 170 C J Fe CuNi thermoelectric sensor thermocouple type J 40 to 800 C K NiCr NiAl th...

Page 10: ...recorder by accident or by unauthorized persons 10 USE AND FUNCTIONS OF THE SD MMC CARD Use of an SD MMC card may be particularly suitable to transfer data in the event of stationary wall installation...

Page 11: ...d in text editors Windows WordPad Notepad etc NOTE If the SD MMC card is installed or removed in the course of recording a new csv file is created where the sequential numbers of events are continued...

Page 12: ...he ARSOFT CFG WZ1 software USB communication is present when mass memory is supported internal memory or the SD MMC card and during operation of the ARSOFT CFG WZ1 software keep a copy of the current...

Reviews: