13. IMPORTANT COMMENTS PERTAINING TO OPERATION  

!

 

Please read this instruction carefully in order to ensure problem-free and optimized operation of the recorder.  

In order to ensure the longest possible operating time of the battery without the need to change it, make sure to:

 

 

set the longest possible value of the measurement and recording interval (recommended > 10 min., 3:

MPer

Table 9.2)

 

 

use an SD/MMC card only when this is actually necessary; otherwise, use the internal memory

 

 

delete unnecessary files from the internal memory or the SD card before new recording starts

 

If one of the following conditions is met:

 

 

the measurement and recording interval is shorter than 10 min.;

 

 

an SD/MMC card is used

 

consider using an accessory power supply unit, which may extend the life of the battery to approx. 8 years. 
Moreover, it is recommended to comply with the following advice: 

 

do not disconnect the device from the computer in the course of communication via the USB interface, 
which is indicated by the “

READ/WRITE”

 and in the ARSOFT-CFG WZ1 software. USB communication is 

present when mass memory is supported (internal memory or the SD/MMC card)  and during operation of 
the ARSOFT-CFG WZ1 software. 

 

keep a copy of the current configuration file (

AR233.cfg

) in external memory (SD/MMC card, computer disk, 

etc.) for the purpose of archiving and future off-line configuration. 

If the recorder does not respond, the battery can be taken out for a short time and its charging level can be 
checked (if the voltage is less than 3.3 V, the battery should be replaced with a new one). 

 

14. CHANGING THE BATTERY 

In order to change the battery, use a flat screwdriver and perform the following actions:

 

 

remove the back wall of the housing (as shown in Figure 14.1)

 

 

carefully take out the control board of the recorder from the housing (Figure 14.2)

 

 

replace the battery with a new one of appropriate type (chapter 4), observing the polarity in the holder 

 

reassemble the device by putting the board and the back wall in their places 

NOTE:  

!

 

When the battery is replaced, the real time clock (RTC) is reset to zero and must be set again using                            
the ARSOFT-CFG WZ1 software; other elements remain unchanged. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
                         

Fig. 14.1. Removal of the back wall 

               Fig. 14.2. Removal of the control board

 

Summary of Contents for AR233

Page 1: ...USER INSTRUCTION DATA RECORDER AR233 Version 1 0 5 2016 07 07...

Page 2: ...ASUREMENT CONNECTOR DESCRIPTIONAND THE ELECTRICAL CONNECTIONS 6 7 CONNECTING TOA COMPUTERAND INSTALLATION OF DRIVERS 6 8 INSTALLATION OF SOFTWARE 7 9 SETTING OF THE CONFIGURATION PARAMETERS 7 9 1 TIME...

Page 3: ...ity that the user changes the battery long operation time with a new battery up to 5 years depending on the recording interval the presence of an SD card and the operating temperature of the device po...

Page 4: ...850 C Ni100 RTD 3 or 2 wire 50 170 C thermocouple J 40 800 C thermocouple K 40 1200 C thermocouple S 40 1600 C thermocouple B 300 1800 C thermocouple R 40 1600 C thermocouple T 25 350 C thermocouple E...

Page 5: ...lity EMC immunity according to the PN EN 61000 6 2 2002 U standard emissivity according to the PN EN 61000 6 3 2002 U standard 1 operation time depends on the measurement period the presence of an SD...

Page 6: ...e View button in the dialog box select the DRIVERS folder press the Continue button when the warning Installing a device appears press Continue anyway 4 The virtual COM port CDC USB to UART is install...

Page 7: ...or or analog signal must be connected and the configuration must be programmed There are parameter configuration methods 1 Use the USB port and the ARSOFT CFG WZ1 software on line configuration connec...

Page 8: ...d 16 ID 1 identification number ID recorded in the csv archive file Notes 1 parameter accessible only from the level of ARSOFT CFG WZ1 on line configuration 9 1 TIMEAND DATE The current time and date...

Page 9: ...np type of input Pt100 thermoresistance sensor RTD Pt100 200 850 C Pt100 Ni100 thermoresistance sensor RTD Ni100 50 170 C J Fe CuNi thermoelectric sensor thermocouple type J 40 to 800 C K NiCr NiAl th...

Page 10: ...recorder by accident or by unauthorized persons 10 USE AND FUNCTIONS OF THE SD MMC CARD Use of an SD MMC card may be particularly suitable to transfer data in the event of stationary wall installation...

Page 11: ...d in text editors Windows WordPad Notepad etc NOTE If the SD MMC card is installed or removed in the course of recording a new csv file is created where the sequential numbers of events are continued...

Page 12: ...he ARSOFT CFG WZ1 software USB communication is present when mass memory is supported internal memory or the SD MMC card and during operation of the ARSOFT CFG WZ1 software keep a copy of the current...

Reviews: