APAR AR233 User Instruction Download Page 5

5

 

Communication interface for work  
with a computer

 

USB, drivers compatible with Windows XP/Vista/7/8/10 

Measurement and data recording interval 

programmable 10 s to 24 hours

 

Data storage memory 

(non-volatile): 

 - internal 

4MB, FLASH, file system FAT12, recording of up to 80,000 
measurements 

 - external SD/MMC card 
   (connection with an ejector) 

FAT16, FAT32, maximum size 2 GB, recommended size ≤ 
1GB and FAT16 

(recording up to 19 million measurements)

 

Real time clock (RTC) 

quartz, date, time, takes leap years into account) 

Optical signaling

 

2 LED’s: “READ/WRITE,” “Status”

 

Power supply 

(lithium battery) 

3.6 V type AA (R6), 2,450 mAh, (e.g. SAFT LS14500) 

Time of operation with a new battery (1)

 

up to 5 years (in temperatures 20 ÷ 30 °C) 

Rated operating conditions  

-30 to 80 °C, <100% RH (no condensation) 

Operating environment  

air and neutral gases, no dust 

Housing 

(wall mounted, protection rating IP20)

 

material ABS UL94-V0, white 

Enclosure dimensions

 

80 x 80 x 25 mm 

Operating position

 

any 

Weight 

approx. 100 g (with a battery) 

Electromagnetic compatibility (EMC) 

immunity: according to the PN-EN 61000-6-2:2002(U) standard 

emissivity: according to the PN-EN 61000-6-3:2002(U) standard 

(1)  

 operation time depends on the measurement period, the presence of an SD/MMC card, and the ambient temperature: 

 

- 5 years (recording period > 10 min., recording in internal memory, data copied only through the USB, 20 - 30 °C) 

 

- 8 months (measurement period 10 s, internal memory, data copied only through the USB, 20-30 °C) 

 

- 1.5 years (measurement period > 10 min., recording on an SD/MMC card, 20-30 °C) 

         5 months (measurement period 10 s, recording on an SD/MMC card, 20-30 °C) 
- an unused SD/MMC card installed in the socket also reduces the operating time of a battery 
- it takes about 2 min. to transfer the entire content of the internal memory (4 MB) onto an SD/MMC card and it consumes about 1-2 mAh of the battery’s   
   capacity (tested on SanDisk and Kingston memory cards) 
- by using an accessory power supply device, the operating time of a new battery can be extended to about 8 years

 

 
 

5. HOUSING DIMENSIONS, DESCRIPTION OF CONNECTIONS AND EXTERNAL ELEMENTS 

 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

In the case of wall-mounted installation, in order to make mounting holes, the back wall of the housing can be 
uninstalled (chapter 14) and used as a template. 

Summary of Contents for AR233

Page 1: ...USER INSTRUCTION DATA RECORDER AR233 Version 1 0 5 2016 07 07...

Page 2: ...ASUREMENT CONNECTOR DESCRIPTIONAND THE ELECTRICAL CONNECTIONS 6 7 CONNECTING TOA COMPUTERAND INSTALLATION OF DRIVERS 6 8 INSTALLATION OF SOFTWARE 7 9 SETTING OF THE CONFIGURATION PARAMETERS 7 9 1 TIME...

Page 3: ...ity that the user changes the battery long operation time with a new battery up to 5 years depending on the recording interval the presence of an SD card and the operating temperature of the device po...

Page 4: ...850 C Ni100 RTD 3 or 2 wire 50 170 C thermocouple J 40 800 C thermocouple K 40 1200 C thermocouple S 40 1600 C thermocouple B 300 1800 C thermocouple R 40 1600 C thermocouple T 25 350 C thermocouple E...

Page 5: ...lity EMC immunity according to the PN EN 61000 6 2 2002 U standard emissivity according to the PN EN 61000 6 3 2002 U standard 1 operation time depends on the measurement period the presence of an SD...

Page 6: ...e View button in the dialog box select the DRIVERS folder press the Continue button when the warning Installing a device appears press Continue anyway 4 The virtual COM port CDC USB to UART is install...

Page 7: ...or or analog signal must be connected and the configuration must be programmed There are parameter configuration methods 1 Use the USB port and the ARSOFT CFG WZ1 software on line configuration connec...

Page 8: ...d 16 ID 1 identification number ID recorded in the csv archive file Notes 1 parameter accessible only from the level of ARSOFT CFG WZ1 on line configuration 9 1 TIMEAND DATE The current time and date...

Page 9: ...np type of input Pt100 thermoresistance sensor RTD Pt100 200 850 C Pt100 Ni100 thermoresistance sensor RTD Ni100 50 170 C J Fe CuNi thermoelectric sensor thermocouple type J 40 to 800 C K NiCr NiAl th...

Page 10: ...recorder by accident or by unauthorized persons 10 USE AND FUNCTIONS OF THE SD MMC CARD Use of an SD MMC card may be particularly suitable to transfer data in the event of stationary wall installation...

Page 11: ...d in text editors Windows WordPad Notepad etc NOTE If the SD MMC card is installed or removed in the course of recording a new csv file is created where the sequential numbers of events are continued...

Page 12: ...he ARSOFT CFG WZ1 software USB communication is present when mass memory is supported internal memory or the SD MMC card and during operation of the ARSOFT CFG WZ1 software keep a copy of the current...

Reviews: