11. VIEWING RECORDED MEASUREMENTS AND EVENTS 

In order to archive data, the recorder creates a text file with the "csv" extension in the 

internal memory or on an 

SD/MMC card. The file name contains the device type (AR233), the identification number 16:

ID

, and the date and 

time of creation of the file, e.g. „AR233_1_2009-11-30_10-57-16.csv”. The format of a single data record is the 
following: “sequential number of the event;date;time;identifier of the event;measurement;internal 
temperature;checksum”, An example record for measurement of 2 channels: “30;2009-11-
09;16:34:58;5;149,5;26,2;8BE2”, where the measurement =149,5 and the internal temperature=26,2 °C. The types 
and the identifiers of the recorded events are:

 

 

measurement (identifier of event 

5

)

 

 

connection to the USB port (

0

, “USB;CONNECTED”)

 

 

disconnection from the USB port (

1

, “USB;DISCONNEC”)

 

 

loading of new configuration parameters (

3

, “NEW;ON-LINE” or “NEW;OFF-LINE” )

 

 

creation of a new “

csv

” file (

4

, “ID;xxxx”, where xxxx - value of parameter 16:

ID

 

of the device)

 

 

low battery voltage, below 3.35 V (

6

, “LVBAT;x.xx”, where x.xx - battery voltage [V])

 

To make a presentation of recorded results in a graphic form or as a text, and to print them, the ARSOFT-WZ3 
software must be used; the software also enabled detecting unauthorized modifications of the archive. As an 
alternative, “csv” files can be edited in spreadsheet software (OpenOffice Calc, Microsoft Excel, etc.) and in text 
editors (Windows WordPad, +, etc.). 

NOTE:  

!

 

If the SD/MMC card is installed or removed in the course of recording, a new "csv" file is created where the 
sequential numbers of events are continued from the previous file. 
 
 

12. MESSAGE AND ERROR SIGNALING 

a) measurement errors recorded in the “csv” archive file in the measured value field: 

Code

 

Possible causes of error 

19999

 

 

the measurement range of the sensor is exceeded from the top 

 

the sensor is broken 

 

the sensor that is connected is different than the one that is set in the configuration (chapter 9.3, 
parameter 5:

inp

)

 

-19999

 

 

the measurement range of the sensor is exceeded from the bottom 

 

the sensor is broken 

 

the sensor that is connected is different than the one that is set in the configuration (chapter 9.3, 
parameter 5:

inp

)

 

b) other elements indicated by the LED’s on the side of the housing  

LED

 

Description of message 

READ/
WRITE

 

 

writing to/reading from the internal memory or the SD/MMC card (also during recording of a 
measurement) 

 

recording is active (a short flash every 10 seconds) 

STATUS

 

transmission through the virtual COM port (used in ARSOFT-CFG WZ1, item 1 of chapter 9) 

copying of the configuration file from the SD/MMC card to the recorder and transfer of files with 
recorded data onto the card after the 

“COPY”

 button is pressed (chapter 10 

full memory that prevents saving recorded data in an archive file - a quick flash during attempted 
recording of measurement 

too low battery voltage (less than 3.35 V) - short flash every 10 seconds when recording is active (also 
a record is created in the archive file, chapter 11)

 

 

  

Summary of Contents for AR233

Page 1: ...USER INSTRUCTION DATA RECORDER AR233 Version 1 0 5 2016 07 07...

Page 2: ...ASUREMENT CONNECTOR DESCRIPTIONAND THE ELECTRICAL CONNECTIONS 6 7 CONNECTING TOA COMPUTERAND INSTALLATION OF DRIVERS 6 8 INSTALLATION OF SOFTWARE 7 9 SETTING OF THE CONFIGURATION PARAMETERS 7 9 1 TIME...

Page 3: ...ity that the user changes the battery long operation time with a new battery up to 5 years depending on the recording interval the presence of an SD card and the operating temperature of the device po...

Page 4: ...850 C Ni100 RTD 3 or 2 wire 50 170 C thermocouple J 40 800 C thermocouple K 40 1200 C thermocouple S 40 1600 C thermocouple B 300 1800 C thermocouple R 40 1600 C thermocouple T 25 350 C thermocouple E...

Page 5: ...lity EMC immunity according to the PN EN 61000 6 2 2002 U standard emissivity according to the PN EN 61000 6 3 2002 U standard 1 operation time depends on the measurement period the presence of an SD...

Page 6: ...e View button in the dialog box select the DRIVERS folder press the Continue button when the warning Installing a device appears press Continue anyway 4 The virtual COM port CDC USB to UART is install...

Page 7: ...or or analog signal must be connected and the configuration must be programmed There are parameter configuration methods 1 Use the USB port and the ARSOFT CFG WZ1 software on line configuration connec...

Page 8: ...d 16 ID 1 identification number ID recorded in the csv archive file Notes 1 parameter accessible only from the level of ARSOFT CFG WZ1 on line configuration 9 1 TIMEAND DATE The current time and date...

Page 9: ...np type of input Pt100 thermoresistance sensor RTD Pt100 200 850 C Pt100 Ni100 thermoresistance sensor RTD Ni100 50 170 C J Fe CuNi thermoelectric sensor thermocouple type J 40 to 800 C K NiCr NiAl th...

Page 10: ...recorder by accident or by unauthorized persons 10 USE AND FUNCTIONS OF THE SD MMC CARD Use of an SD MMC card may be particularly suitable to transfer data in the event of stationary wall installation...

Page 11: ...d in text editors Windows WordPad Notepad etc NOTE If the SD MMC card is installed or removed in the course of recording a new csv file is created where the sequential numbers of events are continued...

Page 12: ...he ARSOFT CFG WZ1 software USB communication is present when mass memory is supported internal memory or the SD MMC card and during operation of the ARSOFT CFG WZ1 software keep a copy of the current...