4. Notice on use and safety warning
*********************
(1) Read this manual carefully before installing and using of the device.
(2) Be sure the device reliable ground connected during use, otherwise may
cause bar body burn.
(3) Touching of electrode tip during use strictly prohibited.
(4) Do not use this device in the environment where the humidity is greater
than 75%.
(5) Input air must be clean and filled air without water, oil and dust particle
(6) It is strictly prohibited to use this equipment in combustible and explosive
environment.
(7) Internal maintenance and repair must be carried out by professionals
5. Installation guide
**********************************
(1) Installation
a. Choose optimal location for discharge and fix the bar and power supply
steadily.
b. Connect earth to earth connect end on ion bar.
c. Insert one end of the power cord into power adapter and the other end into
power port.
d. Connect air connector to air supply device with tube and turn on the air
switch.
e. Turn on the power switch and adjust proper air pressure.
(2) Outline dimensional drawing
POWER
POS
NEG
40
50
6
3
.
5
35
7
8
.
5
24
5.1 8
4
8
28
⑥
⑤
⑧
⑦
①
②
③ ④
①
Power outlet
②
Add
③
Power alarm
④
HV alarm
⑤
Bar body
⑥
Installation base
⑦
Discharge electrode socket
⑧
End cover
(3) Positioning
a. Caution
b. Ion bar should be installed within the working area of static removing
and installation angle should be perpendicular to the surface of the
charged body
c. The discharge electrode of ion bar should stay away from metal
conductor, metal grounding, rods must be reliable ground connected.
d. Trouble shooting
N
O
Error
Possible reason
Solution
1
Power port indication
LED off
Poor contact of power
cord
Check status and stable connection
of power cord
Power supply mismatch
Check power supply specification
(
INPUT
:
100—240VAC
50/60Hz
;
OUTPUT
:
24VDC
1000mA
)