Calibrated sensors
C
NTA47265-300A2
141 / 164
Sensors declared in this manner are added to the list of couple types (or of probe types) proposed in the measurement
function settings dialog box. They appear at the top of the list, in front of the standard sensors. Their name is preceded by
the '*' character, indicating that these are calibrated sensors.
C.5 Calibration procedure
The CALYS 1000 / 1200 is capable of creating a calibration report from a pre-defined procedure.
The number of procedures that can be recorded depends on the size of the available memory and the size of each
procedure (number of test points). If the memory is not being used by other functions, it is possible to record several tens
of procedures.
To find the available memory size, refer to the chapter, Storage of current acquisitions (chapter 6).
Open the MEASUREMENT CONFIGURATION dialog box,
Select the
Calibration procedure
function,
Press ENTER.
Use the Up
/ Down
arrow buttons to adjust the parameters or press the function button
F1
(new) to define a
new procedure,
Press ENTER.
Complete the information fields for the transmitter to be calibrated and define the input/output types and the full-
scale, where appropriate. To move from field to field, use the function button, F1 (
).
Define the calibration points by pressing F2 (Points).