GETTING STARTED
118 / 164 NTA47265-300A2
CALYS 1000 / 1200
Pressure measurement and detection of contact opening
B.2.11
The CALYS 1000 / 1200 can detect the opening or closing of a contact (switch) and acquire the value of pressure at the
moment that this change of state occurs.
In this case, the pressure measurement function is configured from the transmission part. The transmission/simulation
functions are selected by pressing the function button,
F2
(OUT). Pressing this button activates the
transmission/simulation window (see next chapter for more information on the transmission/simulation part).
Open the MEASUREMENT CONFIGURATION dialog box,
Select the
Pressure
measurement function.
Press ENTER.
To access the
Switch Test
mode, press the function button,
F4
, to select the
Mode
menu and then press ENTER.
Select the type of detection (
Open, Close or Open and Close
) and press ENTER.
Once the type of detection is selected, every change of the selected type is detected and the measured value of pressure
at the time of change is displayed.
To accept (acknowledge) the measurement, press the function button,
F1
(OK).
To exit the
Switch Test
mode, press
ESC
.