Generation / Simulation
B
NTA47265-300A2
125 / 164
Frequency generation
B.3.6
Display the
TRANSMISSION CONFIGURATION
dialogue box:
Select the
Frequency
transmission function, then the range using the function and navigation keys.
Select the
Signal
“Output type”.
Enter the amplitude of the signal between 0 and 20 V.
Confirm with ENTER.
The following ranges are available:
Range
1,000 Hz
10,000 Hz
Resolution
0.01 Hz (1)
0.1 Hz (1)
Max. Amplitude
20 V
20 V
(1) Note that the captured value may differ from the displayed value. The frequency generated is derived from a fixed
frequency by dividing it by a whole number. The displayed value (within the resolution of the display) is, therefore, this
value recalculated with the fastest approximation to the captured value.
The display units may be in Hz or Beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.
Frequency generation for dry contact
B.3.7
Display the
TRANSMISSION CONFIGURATION
dialogue box:
Select the
Frequency
transmission function, then the range using the function and navigation keys.
Select the “Type of Output”
Hard contact
.
Confirm with ENTER.
The following ranges are available:
Range
1,000 Hz
10,000 Hz
Resolution
0.01 Hz (1)
0.1 Hz (1)
Max. Amplitude
20 V
20 V
(1) Note that the captured value may differ from the displayed value. The frequency generated is derived from a fixed
frequency by dividing it by a whole number. The displayed value (within the resolution of the display) is, therefore, this
value recalculated with the fastest approximation to the captured value.
The display units may be in Hz or Beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.