Generation / Simulation
B
NTA47265-300A2
123 / 164
Resistive probe simulation (temperature)
B.3.4
Display the
TRANSMISSION CONFIGURATION
dialogue box:
Select the
Rt
transmission function, then the appropriate “type of probe”, and range using the function and
navigation keys.
Select the measurement current
Select the display unit
Confirm with ENTER.
Connection is made between the two
terminals.
The following probes are available:
Sensor
Pt 50 (
= 3851)
Pt 100 (
= 3851)
Pt 100 (
= 3916)
Pt 100 (
= 3926)
Pt 200 (
= 3851)
Pt 500 (
= 3851)
Pt 1000 (
= 3851)
Ni 100 (
= 618)
Ni 120 (
= 672)
Ni 1 000 (
= 618)
Cu 10 (
= 427)
Cu 50 (
= 428)
The resistive temperature probe simulation function can be used with either 2, 3 or 4 wire connection.
As for the resistance simulation function, if a polling acquisition system is used, ensure that the transmitter carries out the
measurement at least 1 ms after the current is present.