B46IB01
29
EN
8.3
Storage of Measuring Results
8.3.1 Manual Storage
•
A long push of the "store" key immediately stores the current measuring result in
the memory together with a sample number.
•
"HOLD" is displayed.
•
The sample number starts at 1 and is automatically increased by one with each
new stored measuring value.
•
1024 results can be stored.
8.3.2 Automatic Storage
•
A short push of the "store" key activates the automatic storage procedure.
•
"AUTO" is displayed.
•
The instrument automatically checks the temperature stability of the sample. If
temperature is stable within 0.2 degrees for 10 subsequent measurements, then
the value is automatically stored together with a sample number.
•
A minimum of 15 seconds is required per automatic storage procedure.
8.4
Recalling Stored Results
•
A short push of the "Recall/Delete" key shows the latest stored value on the
display for 10 seconds.
•
If the "Recall/Delete" key is pushed again shortly within this time, subsequently
all stored results can be displayed from the latest to the oldest value.
•
If the key is not pushed within 10 seconds the display is switched back to the
current measuring value.
Information:
•
No storage is possible if the "low battery" sign appears on
the display.
•
Activating the "SET" mode deletes all stored results.
Summary of Contents for DMA 35N
Page 1: ...B46IB01 3 EN DMA 35N Portable Density Meter Firmware Version V2 014 Instruction Manual ...
Page 39: ...B46IB01 41 EN Appendix C CE Declaration of Conformity ...
Page 40: ...42 B46IB01 EN Appendix D EC Type Examination Certificate ...
Page 41: ...B46IB01 43 EN ...
Page 42: ...44 B46IB01 EN ...
Page 43: ...B46IB01 45 DE DMA 35N Tragbares Dichtemessgerät Firmwareversion V2 014 Betriebsanleitung ...
Page 81: ...B46IB01 83 DE Anhang C CE Konformitätserklärung ...
Page 82: ...84 B46IB01 DE Anhang D EG Baumuster Prüfbescheinigung ...
Page 83: ...B46IB01 85 DE ...
Page 84: ...86 B46IB01 DE ...
Page 85: ...B46IB01 87 FR DMA 35N Densimètre portatif Version du logiciel V2 014 Mode d emploi ...
Page 123: ...B46IB01 125 FR Annexe C Déclaration de conformité CE ...
Page 124: ...126 B46IB01 FR Annexe D Certificat d examen de type CE ...
Page 125: ...B46IB01 127 FR ...
Page 126: ...128 B46IB01 FR ...
Page 127: ...B46IB01 129 ES DMA 35N Densímetro Portátil Versión de Firmware V2 014 Manual de Instrucciones ...
Page 165: ...B46IB01 167 ES Apéndice C Declaración de conformidad CE ...
Page 166: ...168 B46IB01 ES Apéndice D Certificado de examen tipo EC ...
Page 167: ...B46IB01 169 ES ...
Page 168: ...170 B46IB01 ES ...