26
B46IB01
EN
8
Measurement
Caution:
•
Before any measurements are performed, check the
resistance of all materials in contact with the sample
•
Materials in contact with the sample are:
- PTFE (pump piston, filling tube)
- Borosilicate glass (measuring cell, pump cylinder)
•
Before performing any measurements, prepare suitable
cleaning agents or solvents.
Information:
•
The sample has to be filled into the measuring cell free of
any bubbles. Bubbles seriously interfere with the
measurement.
Possible sources of bubbles are:
- Leakage at the connection of the filling tube or the screw
plug.
- Gas bubbles contained in the sample.
•
The measuring cell has to be completely filled.
•
The sample temperature should not be lower than the
ambient temperature.
•
Carefully clean the measuring cell after each series of
measurements to avoid deposit of coatings.
•
The displayed temperature value is flashing as long as the
temperature of the filled sample exceeds the temperature
specification of the DMA 35
N
(0 to 40 °C). Measuring
results will be displayed, but it is recommended to wait
until the temperature display stops flashing.
•
The flashing temperature unit (°C or °F) indicates the
proper functioning of the instrument.
Summary of Contents for DMA 35N
Page 1: ...B46IB01 3 EN DMA 35N Portable Density Meter Firmware Version V2 014 Instruction Manual ...
Page 39: ...B46IB01 41 EN Appendix C CE Declaration of Conformity ...
Page 40: ...42 B46IB01 EN Appendix D EC Type Examination Certificate ...
Page 41: ...B46IB01 43 EN ...
Page 42: ...44 B46IB01 EN ...
Page 43: ...B46IB01 45 DE DMA 35N Tragbares Dichtemessgerät Firmwareversion V2 014 Betriebsanleitung ...
Page 81: ...B46IB01 83 DE Anhang C CE Konformitätserklärung ...
Page 82: ...84 B46IB01 DE Anhang D EG Baumuster Prüfbescheinigung ...
Page 83: ...B46IB01 85 DE ...
Page 84: ...86 B46IB01 DE ...
Page 85: ...B46IB01 87 FR DMA 35N Densimètre portatif Version du logiciel V2 014 Mode d emploi ...
Page 123: ...B46IB01 125 FR Annexe C Déclaration de conformité CE ...
Page 124: ...126 B46IB01 FR Annexe D Certificat d examen de type CE ...
Page 125: ...B46IB01 127 FR ...
Page 126: ...128 B46IB01 FR ...
Page 127: ...B46IB01 129 ES DMA 35N Densímetro Portátil Versión de Firmware V2 014 Manual de Instrucciones ...
Page 165: ...B46IB01 167 ES Apéndice C Declaración de conformidad CE ...
Page 166: ...168 B46IB01 ES Apéndice D Certificado de examen tipo EC ...
Page 167: ...B46IB01 169 ES ...
Page 168: ...170 B46IB01 ES ...