IMPLANTEO
®
7.1 Language selection
Press the
« settings »
button
(Fig.1, rep.4 or Fig.3, rep.11).
The button
(Fig.9, rep.2)
accesses the language selection screen.
Then select the required fl ag and validate.
7. Adjusting parameters
Fig. 9
This is the description of the screen « confi guration » according
to the description
Fig.9
displayed after pressing the button .
1.
Users or Protocols.
2.
Language selection.
3.
Calibration of used contra-angle.
4.
Adjustment of LED lighting.
5.
Adjustment of sound level.
6.
Advanced settings.
7.
Pedal calibration.
8.
Touch pad calibration.
9.
Save/ transfer of the parameters.
10.
Software update.
11.
Re-initialisation of initial confi guration.
12.
Return to previous screen.
1
7
2
8
3
9
4
10
11
12
5
6
7.2 Calibration of contra-angle in use
To make the control and display of the torque value more accurate, the contra-angle needs to be calibrated.
This provides the motor with information about output and effi ciency of the contra-angle in use.
The test is carried out in 2 main phases :
w
First, the contra-angle is made to rotate with no load.
w
The contra-angle must then be connected to the specifi c calibration module. This module simulates an average
intervention in the mouth. The measurement is therefore made in conditions as close as possible to real operation.
w
Before using the calibration module, it MUST be lubricated using the Anthogyr lubricating spray
Ref.
1930X
equipped with
an adaptor
Ref.
1933X
.
w
Place the adaptor over the connection orifi ce.
w
Maintain the spray adaptor in contact on the calibration module.
w
Spray for 1 second.
The calibration module is not sterilizable.
As the capacities of the contra-angles can vary signifi cantly, calibration of the contra-angle is recommended after
maintenance and sterilization, with assistance of a second person to bring the gauge to a standstill (non sterile)
and retrieve the mandrel after calibration.