40
Press the icon
(Fig.2, rep.1 or Fig.3, rep.9)
to access the sequence management screen
(Fig.5)
.
6.2 Adding / Deleting / Replacing sequences
6. Work screen
1.
Flow of peristaltic pump.
2.
Decrease the value.
3.
Adjusted fl ow.
4.
Increase the value.
5.
Setting cursor.
6.
Speed of micromotor.
7.
Set maxi. speed.
8.
Torque limitation of the micromotor.
9.
Maximum reachable torque.
10.
Cancel.
11.
Validate.
6.1 Settings Flow / Speed / Torque
1
6
8
10
11
7
9
2
3
4
5
Fig. 4
This is the description of the screen « settings Flow/ Speed / Torque » according to the representation
Fig.4
:
1.
Active User / Protocol and access to welcome screen.
2.
Reduction ratio for active contra-angle and choice of it.
3.
Activate or deactivate the brightness of the contra-angle.
4.
Instantaneous speed of the tool if the motor is active or speed limit
when stopped.
5.
Instantaneous torque if the motor is active or maxi. reached when
the motor is stopped.
6.
Active sequence of all the sequences included in the active protocol.
7.
Programmed torque limit.
8.
Access to graphical display.
9.
Edition of sequences of the current protocol.
10.
Rotation direction of the motor and reversed direction.
11.
Access settings.
12.
Adjust the fl ow of the peristaltic pump.
13.
Irrigation level.
Tactile zone 13, 4, 5, 12, 7 :
Access settings Flow/ Speed/ Torque.
Select a user protocol and validate
(Fig.1, rep.5)
to access the work screen
(Fig.3)
.
When functioning, only the main data are displayed on the screen : Flow / Torque / Speed.
When switching on the device, the welcome screen only stays active for 10 seconds before toggling to the work screen
(Fig.3)
associated with the active user / protocol.
This is the description of the « work screen » according to the representation
Fig.3
.
C
Anthogyr
25
Rpm
N.cm
35
1
2
3
7
5
4
8
9
6
12
13
10
11
Fig. 3
Press on buttons « + » and « - » or move the cursor on the bargraph.
Press the centre of the screen
(Fig.3)
to have access to the settings of the 3 main data: irrigation fl ow, maximum speed
and maximum torque.
With the work screen, it is also possible to adjust the fl ow of peristaltic pump with a short press on button A of the pedal
(see section
« 8 – Using the pedal »
).
C