viii
User manual OR 100 /110, ed. 5
protective earth via a ground socket. The power source should exhibit min-
imal power transients and fluctuations. Replace faulty or frayed power
cords.
The OR 100/110 is equipped with 8 AC power outlets at the back panel, di-
vided in two blocks of 4. The maximum power, which can be drawn from
these power outlets, is 1900 Watt at 240VAC, 960 Watt at 120V and 800
Watt at 100 VAC. The maximum rated power per outlet is 300 Watt.
Do not exceed the maximum rated power, this can cause fire or other po-
tentially dangerous situations. It is advised to use the outlets only to con-
nect ALEXYS system components, such as the LC 100 pump, AC 100
controller, AS 100 autosampler and DECADE II
™ electrochemical detec-
tor.
In case of the OR 100 (not applicable for OR 110) always have the silicone
tubing connected, which is in the ship kit, to the drain of the mobile phase
tray at the back side of the OR 100, to prevent leakage of solvents onto