Chapter 5 Maintenance
41
degasser and pulse damper. Plug all of the ports on the OR 100 with the
appropriate caps.
Before using the OR 100/110 again, completely purge the degasser and
pulse damper with the correct mobile phase before reconnecting the col-
umn and restarting the system.
Avoid precipitation of buffer salts in organic solvents. Start always
with flushing system with water when switching solvents.
Replace blown fuses with fuses of proper type and rating (8AT/250V,
UL/CSA approved ) stipulated on the side panel and specified in the instal-
lation section of this manual. The fuse holder is integrated in the mains in-
let.
Fig. 14. Fuse holder.
To access the fuse holder, push plastic lever on the bottom side of the
holder completely and pull holder out of inlet. Ensure that the instrument is
never put in operation with fuses of a different type. This can cause fire.