80 DECADE II user manual, ed. 11
Decreased sensitivity (low S/N ratio)
Possible cause
Remedy
Fouled WE by dirty samples Clean WE, if possible: dilute samples
Cell potential too low
Optimise potential
Contaminated buffer (high
I
cell
)
Replace buffer, do not recycle the buffer
Base line oscillations
Possible cause
Remedy
Malfunctioning pump
(regular pattern)
Check pump (seals, valves)
Over-tightened cell bolts
Adjust cell bolts, check pump pressure
Air bubbles in cell or REF
Maintenance REF
Temperature oscillations
Set oven temperature
Contaminated buffer (high
I
cell
)
Replace buffer, do not recycle the buffer
Fouled WE
Clean WE
Fe
++
in buffer
Add EDTA, passivate metal parts with HNO
3
Saturation of output
Possible cause
Remedy
Damaged REF
Check with spare REF, replace if necessary
Damaged WE
Replace cell block
Cell incorrectly connected
Check connections (REF: black, WE: red,
AUX: blue)
Cell potential too high
Optimise cell potential