CHAPTER 10 Error messages
75
C H A P T E R 1 0
Error messages
Table X. Error messages.
Error
Message
01
Incompatible boot version
02
Control board error
03
Sensor board x error (x = board number)
04
Firmware program error
05
Record error
06
Incompatible FW version
07
Incompatible FW
08
Control board FW erase failed
09
Sensor board x
10
Upload checksum error.
11
Checksum error.
12
Temperature sensor 1 error.
13
Disconnect flow cell x
14
Control board SRAM error.
15
Sensor board x SRAM error .
Please contact your local supplier if one of the above errors occur.
Furthermore the following messages can be displayed on the LCD screen
during a measurement:
Table XI. Messages.
Message
Advice
01 Out of range
Output is either above +1.0V or below
–1.0V.
Pressing AZERO may give an adequate read-out
again. If the message remains after pressing
AZERO, the autozero function is unable to
compensate the background cell current. Advice:
use a less sensitive range in the SETUP menu.
02 PAD overload
Charging current in pulse mode out of range.
Pressing AZERO may give an adequate read-out
again. If not, it is advisable to change the pulse
settings (increase t1) or use a less sensitive range.
Summary of Contents for DECADE II 171
Page 12: ...12 DECADE II user manual ed 11...
Page 18: ...18 DECADE II user manual ed 11...
Page 32: ...32 DECADE II user manual ed 11...
Page 48: ...48 DECADE II user manual ed 11...
Page 54: ...54 DECADE II user manual ed 11...
Page 71: ...CHAPTER 8 Optimisation of working potential 71...
Page 72: ...72 DECADE II user manual ed 11...
Page 76: ...76 DECADE II user manual ed 11...