Selecting the Signal Standard and Channel
Before making measurements, the Site Master must be set to the appropriate wireless net-
work air interface standard and channel number. The most common standards and channel
numbers are stored in the Site Master and, when selected, the Site Master automatically se-
lects the correct frequency and span. The frequency and span can also be manually entered
if necessary.
The following procedure sets the signal standard and channel number automatically by se-
lecting from the standards stored in the Site Master:
Procedure
Step 1.
On the Site Master, press the
MODE
key.
Step 2.
Use the Up/Down arrow key to select the Spectrum Analyzer mode and press
ENTER
.
Step 3.
Select the
Signal Standard
soft key, use the Up/Down arrow key to select a sig-
nal standard, and press
ENTER
.
Step 4.
Select the
Select Channel
soft key and use the Up/Down arrow key or the nu-
meric keypad to enter a valid channel number. Press
ENTER
to accept the
value, or
ESCAPE
to restore the previous value.
The currently selected Center Frequency, Span, Signal Standard, and Channel are displayed
at the bottom of the status window.
5-8
Chapter 5 Spectrum Analyzer Measurements