Step 3.
Press “Next” for the next prompt (below). The 3653 or 3753LF
Calibration Kits contain a disk with calibration data. The
data has to be installed from the MS462C front panel. To do
so, place the floppy disk from the kit into the MS462XC disk
drive. Press the
Cal
key then
COMPONENTS UTILITIES/IN-
STALL KIT INFO FROM FLOPPY DISK.
softkeys
.
Step 4.
Press “OK” for the next step (below). Connect the Open to
Port 2.
Step 5.
Press “OK” for the next step (below). Connect the Short to
Port 2.
OPERATIONS, CALIBRATION
HOT S22
ME7840A OMM
4-9