NOTE
The linear power calibration takes up-to-5 minutes to
complete. It calibrates the output power at the refer-
ence plane for both internal sources (power out of
Port 1 then Port 3) according to the target power
across the frequency range set in the calibration speci-
fication file. When the test completes, the “Next” but-
ton will become available.
Step 5.
Press “Next” to calibrate flat power response (below).
NOTE
The flat power calibration also takes less than 1 min-
ute to complete. It calibrates the output power at the
reference plane for both internal sources (power out
of Port 1 then Port 3) according to the target power
across the frequency range set in the calibration speci-
fication file. When the test completes, the “Next” but-
ton will become available.
POWER LEVEL
OPERATIONS, CALIBRATION
4-6
ME 7840A OMM