Appendix B — Error Messages
B-1
Introduction
This chapter provides a list of information and error messages that could be displayed on the
MT8222A BTS Master. If any error condition persists, contact your local
(http://www.anritsu.com/Contact.asp).
B-2
Self Test or Application Self Test Error Messages
Overall Status FAILED
One or more elements of the System or Application Self Test has failed. Refer to the other
pass fail tests listed below to determine which specific test failed.
ADC Self Test FAILED
The Analog to Digital converter failed to return an answer. Insure that the battery level is
adequate for operation or that the temperature is within acceptable limits. Reset to factory
defaults with either Factory Defaults, ESC+ON, or MASTER RESET, ON. Caution:
Use of MASTER RESET, ON, will erase all user saved setups and measurement
traces and return the unit to a fully Factory Default condition. If the error persists, contact
your Anritsu Service Center.
DDC FAILED
The Digital Down Converter failed to return a value. Insure that the battery level is adequate
for operation or that temperature is within acceptable limits. Reset to factory defaults with
either Factory Defaults, ESC+ON, or MASTER RESET, ON. Caution: Use of
MASTER RESET, ON, will erase all user saved setups and measurement traces and
return the unit to a fully Factory Default condition. If the error persists, contact your Anritsu
Service Center.
Lock Test FAILED
One or more Phase Lock Loops Failed to properly achieve Lock Status. Insure that the
battery level is adequate for operation or that temperature is within acceptable limits. Reset
to factory defaults with either Factory Defaults, ESC+ON, or MASTER RESET, ON.
Caution: Use of MASTER RESET, ON, will erase all user saved setups and
measurement traces and return the unit to a fully Factory Default condition. If the error
persists, contact your Anritsu Service Center.
Over Power FAILED
RF Power applied to the input connector is too high. Remove or reduce the input power or add
additional attenuation. Sometimes out of band frequencies may be present that can cause an
Over Power Error. In highly rich RF environments it may be necessary to add an external
band pass filter to reduce unwanted interference. See the accessories section for a list of
available band pass filters from Anritsu. Out of band frequencies can often be detected by
increasing the Span to maximum in the peak detect mode of operation. Another resolution
may be to reset to factory defaults with either Factory Defaults, ESC+ON, or MASTER
RESET, ON. Caution: Use of MASTER RESET, ON, will erase all user
saved setups and measurement traces and return the unit to a fully Factory Default
condition. If the error persists, contact your Anritsu Service Center.
Summary of Contents for BTS Master MT8222A
Page 2: ...BTS Master User Guide MT8222A The High Performance Handheld Base Station Analyzer...
Page 6: ......
Page 7: ......
Page 9: ......
Page 49: ...3 8 PN 10580 00156 Rev T BTS Master UG...
Page 71: ...5 10 PN 10580 00156 Rev T BTS Master UG...
Page 83: ...7 8 PN 10580 00156 Rev T BTS Master UG...
Page 109: ...C 8 PN 10580 00156 Rev T BTS Master UG...