Using the 2300-580-R Software with VectorStar ME7838D4
3-6 Application Interface – Setup Menu Tab
3659 0.8 mm Cal/Ver Kit & 2300-580-R PVS UG
PN: 10410-00327 Rev. D
3-13
3-6
Application Interface – Setup Menu Tab
The
PVS Application Interface
dialog box displays three tabs with configuration, control, and serial number
information. Each tabbed dialog box contains buttons to control program operations as outlined in the
following sections:
•
The
Setup Menu
tab is described in this section.
•
The
Main Menu
tab display is described in
Section 3-9, “Application Interface – Main Menu Tab”
.
•
The
Serial Number Info
tab is described in
Section 3-10, “Application Interface – Serial Number Tab
.
Procedure
The following procedure continues the program setup from the previous section:
1.
In the
Calibration Kit Type
area, select
0.8 mm Coax
.
2.
In the
Serial Number (5 Digits)
area, enter the serial number for the 3659 Calibration/Verification Kit. At
least five (5) digits are required.
3.
In the
Load Cal-Kit Coefficients to VNA
area, select
Cal-Kit
.
Figure 3-20.
ME7838D4 Application Interface – Setup Menu Tab