12
7.Abnormal state
display
LCD displays “L
℃
” if the measured temperature is
below the minimum of measurement range.
LCD displays “H
℃
” if the measured temperature is
below the minimum of measurement range.
LCD displays “Er1”if the temperature measurement
circuit is abnormal (fault of SENSOR or the
temperature measurement circuit), or the ambient
temperature exceed 10
℃
~40
℃
(50
℉
-104
℉
).
There shall be “ErC” if the calibration process is not
completed or EEPROM is abnormal.
8.Sound
volume
≧
50 db (the perpendicular distance from dB
Volume sensor to thermometer is 10cm )
9.Automatic
shutdown function 10s±1s
10.Low-voltage
display function
The product shall display low-voltage signal if the
voltage is below 2.51V±0.15V.
11.Memory function Memorize 20 groups of measured temperature.
12.Current
consumption
I
stand-by
<2μA; I
working
<0.5mA; I
Buzzer on
<2mA
;
I
banklight
<15mA
13.LED backlit
specifications
≧
1.2cd/m
2
15.Operational
conditions
ASTM
10
℃
-40
℃
(50
℉
-104
℉
)/15-95%RH
16.Type of
measuring
Applicable for forehead temperature and ear
temperature
17.battery
AAA×2
18.Battery life
More than 3000 times
19.Accuracy for
clinical test
The maximum allowable error for clinical test is
specified in the formula below:
≦
0.3
℃
/0.6
℉
(for 95%)
Among which:
T1 and T2 represent temperature value for
thermometer under test respectively, Tref
represents the constant reference temperature
The date of
production
See the label
Terror
=
T1-Tref
T2-Tref
+
2