11
Symptom
Possible Cause
Solution
than the typical
body temperature
range.
dirty.
The distance between
the temperature probe
and the target is too
long.
Move the thermometer
closer to the target.
The thermometer is not
used within 30 minutes
after being taken from a
cold environment.
Wait for more than 30
minutes after the
thermometer is moved
into the measurement
environment.
The temperature
reading is higher
than the typical
body temperature
range.
The temperature probe
is faulty.
Contact ANKOVO.
10. Technical specifications
Items
Standards
models
Dual mode infrared thermometer JPD-FR100+
1.Applicable
regulations and
laws
ASTM 1965/EN12470-5
2.Temperature units
℃
/
℉
, adjustable
3.Measurement
range
Forehead temperature mode:32.0
℃
-42.2
℃
/
89.6
F – 107.9
F
ear temperature mode:0.0
℃
-100.0
℃
/32.0
F –
212.0
F
4.Precision
±0.2
℃
/±0.4
F
5.Display resolution 0.1
℃
/0.1
℉
6.Latency Time
1 second