15
JPD-FR100+ Infrared Thermometer should assure that it is used in such an
environment.
Emissions
test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The JPD-FR100+ Infrared Thermometer
uses RF energy only for its internal
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The JPD-FR100+ Infrared Thermometer is
suitabl e for use in all establishments,
including domestic establish ments and
those directly connected to the public
low-voltage powe r supply network that
supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration – electromagnetic
immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The JPD-FR100+ Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or the user of the
JPD-FR100+ Infrared Thermometer should assure that it is used in such an
environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic
material,
the
relative humidity should be
at least 30 %.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.