Operator’s Manual
pg. 32
Rev 4/24/20
If you see a leak
If you see a leak, execute the Initial Maintenance procedure above, tightening the Packing Nut 1/16 of a turn. If the
leak continues, replace the packing and agitator.
Replace the Fuses (as needed)
To replace the fuses in the ANKOM
DELTA
Automated Fiber Analyzer, follow the procedure detailed below.
1.
Turn off the instrument power and unplug the
Power Cord from the outlet.
2.
Push in and twist the black cap
counterclockwise ¼” turn to release the fuse.
3.
Check both fuses.
4.
Pull the fuse from the black fuse cap and replace
as needed.
120V – 15 amp (Part#687) / 220V – 10 amp
(Part#6035)
Clean the Fiber Optic Level Sensor
Using a cotton swab with alcohol, wipe the tip of the level
sensor at least once per month if you use the instrument
daily.
Level Sensor
Fuses (2)