Operator’s Manual
Rev 4/24/20
pg. 13
ADF Analysis
ADF Calculation
ADF contained within a food or feed sample can be calculated using the following formula:
% ADF
=
100 x (W
3
–
(W
1
x C
1
))
(as-received basis)
W
2
Where:
W
1
=
Bag tare weight
W
2
=
Sample weight
W
3
=
Dried weight of filter bag with fiber after extraction process
C
1
=
Blank bag correction (running average of final oven-dried weight
divided by original blank bag weight)
ADF Sample Preparation Procedure
To prepare samples for fiber analysis, follow the procedure detailed below.
IMPORTANT:
When using the ANKOM
DELTA
Automated Fiber Analyzer for ADF analysis, at
least one blank filter bag should be included with the sample set as an indicator
of particle loss. A running average of the blank bag weights is used in the fiber
calculation as the C
1
correction factor. A C
1
value larger than 1.0000 indicates
that sample particles were lost from filter bags and deposited on the blank bag.
Any fiber particle loss from the filter bags will generate erroneous results. If
particle loss is observed, the grinding method for the specific sample should be
evaluated.
1.
Using a Solvent Resistant Marker, number all of the filter bags you will use during
the fiber analysis.
2.
Weigh and record the weight of each empty filter bag (W
1
).
3.
Set the Heat Sealer dial to between 4 and 5. (The setting may vary from sealer to
sealer.)
4.
Seal at least one empty filter bag (to be used as a blank) within
4mm of its open end. Keep the sealer arm down for 2 – 3
seconds after the red sealer light turns off (to cool the seal). The
seal can be seen as a solid melted stripe along the top edge of
the filter bag (as shown to the right). If the seal is not strong, re-
seal the bag.
5.
Place an empty filter bag in the Bag Weigh Holder in an open position.
6.
Tare the weight of the empty filter bag and the holder together.
7.
Add 0.45 – 0.50g of sample to the filter bag. Keep all particles away from the sealing
area of the filter bag.
8.
Record the weight of the sample (W
2
).
Seal
Summary of Contents for Delta
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 18: ...Operator s Manual pg 18 Rev 4 24 20 This page intentionally left blank ...
Page 24: ...Operator s Manual pg 24 Rev 4 24 20 This page intentionally left blank ...
Page 30: ...Operator s Manual pg 30 Rev 4 24 20 This page intentionally left blank ...