Flush Procedure
The Flush procedure allows you to clean the system with water. The instrument automatically flushes the system after
each run however if a user would like to perform an additional flush procedure, they can do so in diagnostic mode.
To perform a Flush of your analyzer, follow the instructions below.
1.
Turn the instrument’s Power Switch to the ON position. The Display will light.
2.
Verify that the water supply is on and the drain hose is securely in the drain.
3.
Press the arrow keys on the Keypad until you see “Diag Mode”
on the Display. Press Enter.
4.
Scroll to the desired valve and click Enter to open the valve.
5.
Press and hold START, water will flow through the valve.
NOTE:
If supply line or amylase container is attached to the inlet side of the valve, water will
flow into the supply line (or amylase container). Flushing the valve for 1-2 seconds
is usually sufficient.
Diag Mode
v
<Enter>