Environmental Conditions:
Operating Conditions:
Temperature of 15°C to 35°C.
Humidity of 20% to 80% non-condensing.
Height above sea level to be less than 2000m.
Movement and Storage Between Use:
Temperature of 5°C to 40°C.
Humidity of ≤ 80% non-condensing.
Atmospheric Pressure 50kPa - 113kPa.
Floor to be level to within 10° of horizontal when being moved.
Not suitable for negotiating steps or thresholds.
Initial Transport and Storage in Original Packaging:
Temperature of -20°C to 40°C.
Humidity of 20% - 90% non-condensing.
Atmospheric Pressure 50kPa - 113kPa.
EMC Data:
Electromagnetic Disturbances;
The AT4 Tourniquet is compliant to EN IEC 60601-1-2:2015 the “Collateral standard for
Electromagnetic disturbances - Requirements and tests” and is accordingly classified as
a Class A device for professional use.
To reduce the risks of electromagnetic interference best practice as detailed below
should be observed.
If it is necessary to use the AT4 Tourniquet in close proximity to other electrical
equipment, observe the performance of both the AT4 Tourniquet and the other
equipment to make sure that they are operating normally. Ensure cables are not
coiled or bundled in close proximity to each other and are separated from other
cables where possible.
The AT4 Tourniquet is suitable for use in hospital operating rooms and other
clinical areas with antistatic flooring. The emissions characteristics of this
equipment make it suitable for use in hospitals, it is not intended for use in a
residential environment.
To reduce risks of interference electromedical equipment should
not be placed in close proximity to radio frequency communication
equipment or electrosurgical equipment. Interference may occur in
the vicinity of equipment marked with the symbol indicated.
If interference occurs the user may need to take mitigation measures, such as
moving or reorienting the equipment or separating cables.
9.
Cleaning and Disinfecting
The AT4, and its detachable hoses/cables must
NOT
be immersed in water or other liquids
during cleaning or disinfection. Do
NOT
use solvents or abrasive cleaners
Cleansers and disinfectants must be CE marked indicating an intended purpose of medical
devices & specified for use on plastics and metal surfaces. Suitable disinfectants include:
quaternary ammonium compounds, isopropyl alcohol, chlorine or chlorine dioxide 0.5%
and phenolics.
CAUTION: Before cleaning, disconnect from the mains electrical supply.
Wipe the AT4 and its detachable hoses/cables using a cloth dampened with detergent
diluted with water as per the manufacturer’s instructions. Apply the liquid to the cloth
and squeeze out surplus liquid. Do not apply liquid directly to the AT4 or its detachable
parts.