11.

 

Expected battery life 

When  running  on  DC  power  only,  a  fully  charged  unit  should  be  able  to  perform  the 
following number of bi-lateral 30 minute procedures at 300mmHg; 
 

 

35* times with battery indicator displaying green 

 

A further 15* times with battery indicator displaying amber 

 

Once  the  battery  indicator  begins  displaying  red  the  device  should  only  be  used  while 
remaining connected to the mains supply. 

 

* Based on well-maintained battery. 

12.

 

Maintenance 

12.1.

 

Daily 

Ensure that O rings, cuffs and associated hoses are in good condition  before use. Those 
that are damaged or worn should be replaced. 
 

 

CAUTION:  Before  use,  ensure  the  device’s  functions  operate  correctly.  Also 

visually  inspect  the  device  for  any  loose  or  damaged  parts.  If  the  device’s 
performance or mode of operation  changes from that specified or required 
the  device  should  be  taken  out  of  service  immediately.  If  after  checking  O 
rings,  cuffs  and  associated  hoses  are  in  good  condition  the  AT4  is  still  not 
operating as expected then request maintenance before returning the device 
to clinical use. 

 

If there is any sign of damage or a change in performance the device should be taken out 
of clinical use and maintenance requested.  

 

If the battery indicator of the electrically powered AT4 shows amber connect 

it to the mains electrical supply as soon as possible. If the indicator shows red connect to 
the mains electrical supply immediately. Green shows normal operating conditions 

 

The spanner symbol with an amber or red indicator requires investigation. If 
the indicator is not cleared by replacing leaking cuffs hoses and O rings and 

then restarting the AT4 a service should be requested 

12.2.

 

Recurrent (Periodic) Testing 

 

 

CAUTION:  In  line  with  the  MHRA  Device  Bulletin  DB2006(5),  maintenance 

work  should  only  be  conducted  by  suitably  trained  personnel  following 
manufacturer’s guidelines and approved parts. 

 

It  is  recommended  that  the  device  is  serviced,  electrically  safety  tested  and  the 
performance  and  accuracy  confirmed  on  an  annual  basis  in  accordance  with  the 
manufacturer’s service schedule and EN IEC 62353:2007. 
The electrically powered AT4 contains a lead acid battery pack which will require to be 
replaced every two years. The device is class I electrical safety with a protective earth, it is 
NOT classified as an applied part. 

 

When requesting service ensure that both the Ref No and the Serial No are quoted. The 
Serial and Reference Numbers are located on a label on the rear of the device. 

13.

 

Product Warranty 

The  product,  when  new,  is  guaranteed  to  be  free  from  defects  in  materials  and 
workmanship and to perform in accordance with the manufacturer’s specification for a 

Summary of Contents for 40080

Page 1: ......

Page 2: ......

Page 3: ...Warnings Cautions and Side effects 2 3 Symbols 3 4 Getting Started 5 4 1 Fitting the Fuse and Charging the Battery 5 4 2 Connecting and Storing the Cuff Hoses 5 5 Product Functions 6 6 Controls and O...

Page 4: ...is intended for the control of pneumatic tourniquet cuffs in operating theatres or similar environments 1 2 Intended user Tourniquets should only be operated by trained and competent clinical staff a...

Page 5: ...n accordance with the manufacturer s guidelines see Section 9 CAUTIONS Before use ensure all device functions operate correctly Also visually inspect the device for any loose or damaged parts If the d...

Page 6: ...ational Red indicates incorrect action TIMER REMINDER Prior to inflation of cuff this button sets the reminder start time and repeater frequency This function is cancelled after the cuff has been infl...

Page 7: ...TION Maximum power consumption in watts SN SERIAL NUMBER Unique serial number used for traceability REF REFERENCE NUMBER Reference or model number indicating the type of unit DATE OF MANUFACTURE MANUF...

Page 8: ...y the fuse symbol on the underside of the AT4 On receipt or after periods of storage the AT4 must be connected to the mains electricity supply with the cable provided for 24 hours to allow the battery...

Page 9: ...5 Product Functions 1 Control Panel 2 Cuff Supply Hose Storage Connectors 3 Cuff Supply Hose Connectors 4 Cuff Supply Hose 5 Pulling Handle 6 Cuff Hooks 7 Storage Facility 8 Additional Storage Facilit...

Page 10: ...charged and the battery indicator is green if not it must be used connected to the mains supply The red and blue cuff hoses should be connected to the connectors in the front of the AT4 ready for use...

Page 11: ...rotating the control clockwise to increase and anticlockwise to decrease The selected pressure in mmHg is displayed in the window above the rotary control Application of a tourniquet cuff at excessive...

Page 12: ...associated with the pressure set in the previous step If this button is depressed prior to setting the pressure levels above the device will lock out To reset turn the IVRA button off and then back on...

Page 13: ...e Battery Charging Section 9 When the electronic unit is being used while connected to the mains supply Green Amber Alternating Green Amber indicating connection to the mains while in use NOTE The amb...

Page 14: ...result in damage 8 Performance and Technical Specification Accuracy and Resolution Pressure is measured to an accuracy of 2 5 mmHg and displayed with a resolution of 5 mmHg Maximum Cuff Pressure The...

Page 15: ...ssible The AT4 Tourniquet is suitable for use in hospital operating rooms and other clinical areas with antistatic flooring The emissions characteristics of this equipment make it suitable for use in...

Page 16: ...t is recommended that when not in use the device is connected to the mains electrical supply to recharge the battery The battery will not be damaged by leaving it connected to the mains when fully cha...

Page 17: ...indicator of the electrically powered AT4 shows amber connect it to the mains electrical supply as soon as possible If the indicator shows red connect to the mains electrical supply immediately Green...

Page 18: ...er expressed or implied warranty is given For details of our extended warranty packages please contact Anetic Aid or your authorised dealer 14 Disposal of Waste Electrical Electronic Equipment This sy...

Page 19: ...This page has been left blank MDM 992043 10 120121...

Page 20: ...This page has been left blank...

Reviews: