6/17 

 

 

 

 

 

 

 

 

 

 

 

    Picture 3-1                  Picture 3-2                  Picture 3-3 

 

 

   
 
 
 

 

      Picture 3-4                Picture 3-5                  Picture 3-6 

e.  During  adjusting  clock  and  date,  the  monitor  will  go  back  to  Standby  Mode 

automatically when no button will be pressed within 30 seconds. 

f. 

You can turn off the monitor by pressing “START/STOP” button when the minute 

is blinking, then the time and date is confirmed. 

Note: 

3.1  The clock format could be set by user. 

 

 

4.  CONNECTING THE CUFF TO THE MONITOR 

 

The cuff is attached to the monitor when it is packaged. Should the cuff become 

unattached, align the two plugs and four brackets of the cuff with the plug sockets and 

bracket sockets of the monitor and press the cuff to the monitor until the plugs and 

brackets are securely attached. 

Summary of Contents for KD-723

Page 1: ...OCEDURES 5 1 DOWNLOAD THE FREE APP 5 2 BATTERY LOADING 5 3 CLOCK AND DATE ADJUSTMENT 5 4 CONNECTING THE CUFF TO THE MONITOR 6 5 APPLYING THE CUFF 7 6 BODY POSTURE DURING MEASUREMENT 7 7 TAKING YOUR BL...

Page 2: ...usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and ac...

Page 3: ...ored in the memory with date and time stamp The voice function will ease the operation The Electronic Sphygmomanometers corresponds to the below standards IEC 60601 1 2005 corr 1 2006 corr 2 2007 EN 6...

Page 4: ...0 4 122 15 Environmental humidity for storage and transport 85 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 170 times 18 Wireless Connection Smart Bluetooth Frequency Band 2 400 2...

Page 5: ...formation regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see part E...

Page 6: ...f according to local regulations at the end of their usage 3 CLOCK AND DATE ADJUSTMENT a At first the Blood Pressure Monitor is totally off once you insert the battery the Blood Pressure Monitor will...

Page 7: ...or by pressing START STOP button when the minute is blinking then the time and date is confirmed Note 3 1 The clock format could be set by user 4 CONNECTING THE CUFF TO THE MONITOR The cuff is attache...

Page 8: ...o not move your arm body or the monitor during measurement Stay still calm for 5 minutes before blood pressure measurement Please keep the cuff clean Clean the cuff by wet soft cloth and mild detergen...

Page 9: ...ly inflated The blood pressure and pulse will be measured during inflation Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed on the LCD The irregular...

Page 10: ...Press MEM button again to review the next result See picture 8 6 In this way repeatedly pressing the MEM button displays the respective results measured previously Picture 8 5 Picture 8 6 e When disp...

Page 11: ...e different levels of blood pressure 12 TECHNICAL ALARM DESCRIPTION The monitor will show HI or Lo as technical alarm on LCD with no delay if the determined blood pressure systolic or diastolic is out...

Page 12: ...to detect diastolic pressure LCD shows Er 3 Pneumatic system blocked or cuff is too tight during inflation Apply the cuff correctly and try again LCD shows Er 4 Pneumatic system leakage or cuff is to...

Page 13: ...chnical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measu...

Page 14: ...ANT INFORMATION REQUIRED BY THE R TTE This product is approved in accordance to R TTE directive transmitter interference This product complies with Industry Canada IC RSS 247 The Bluetooth word mark a...

Page 15: ...e KD 723 is intended for use in the electromagnetic environment specified below The customer or the user of the KD 723 should assure that it is used in such an environment IMMUNITY test IEC 60601test...

Page 16: ...NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from struct...

Page 17: ...aximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz...

Page 18: ...rgy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particu...