Instruction Manual
–
V6.19
GYROTORK DUAL version
- 18 -
2.
Output actions
After each calculation, it is possible to activate an output of the torque gauge.
Select the output control from the drop down menu to activate (from 1 to 6) then select its running
mode: pulse or switch mode and the high or low state. For more details on the input/output see
“Chapter X –
Input & Output”.
These actions can be very useful when you want to communicate with
the test stand. Example: stop the test stand when the torque gauge detects the torque break.
VIII.
RECORD DATA AND STATISTICAL Mode.
The GYROTORK has a memory of at least 2.000 values per channel and automatically calculates the
sleeping mean, the standard deviation and the mean divided by the standard deviation. It is possible
to record 3 results simultaneously. Choose between: the maximum, the minimum and calcultaion1
and calcultion2.
Note: The torque
gauge can’t record multiples curves. Only the last curve is stored. You have to
connect the torque gauge to a computer or use the USB memory stick called DATASTICK to save
multiple curves.
Statistics and memory can be enabled by touching the statistics symbol on the setup screen.
1.
Statistics setup
1: Select the desired result to be entered into the statistic arrays; you can choose 3 types of values to
record: maximum, minimum, calculation1 or calculation2 for each channel. Select the units desired.
2: More than one result can be averaged to create a saved result. Enter this number in the Samples.
3
1
2
4