Troubleshooting the LAN Interconnection
Software does not install properly
•
Make sure your memory stick is not damaged.
Software cannot detect any instrument
•
Make sure you have connected both computer and instrument to a common network.
•
If a direct connection is used you may require to reset your computer Ethernet controller (depending
on the configuration). Note that in that case detection of the instrument can take a considerable
amount of time if your computer is configured to work with an external DHCP server. In some cases
the detection may even fail completely. Configure your computer network controller to an
appropriate fixed IP instead.
•
Make sure that your software firewall enables the GUI to setup a TCP/IP connection via the LAN.
Under Windows 7/10 you can do that like this:
Open
Control Panel
under
Settings
in your
Start
menu. Then go to
Windows Firewall
. Click on
Exceptions
and then
add Program.
If the GUI is in this list, choose it and click
OK
otherwise you have
to browse for the path to GUI installation directory. Finally close all open dialogs with
OK
. Now your
Windows™ Firewall will not block requests from the GUI.
Shutting Down the Signal Generator
Press the line on/off switch on the rear panel to off.
Replacing the Fuse
The 2URM case has a fuse, accessible from the outside. To change the fuse, pull out the mains plug and pull
the fuse holder out. Replace the old fuse with a new one. It is forbidden to repair defect fuses or to bridge
them by any means. Use only a fuse with same specifications.