START-UP SEQUENCE
When the C16 power switch is activated, a number of messages are sequentially
displayed. Note that the display always has a row of text across the bottom. The four switches
on the front panel are called “soft switches” because they have different functions depending on
what mode of operation the unit is in. The text just above each switch indicates what that switch
will do when pressed. The following is the display sequence at startup. The sample displays use
ammonia as an example, but the gas symbols, range, and setpoint information will depend on
what sensor module is installed. Pressing the NEXT key will step to the next display. Pressing
the SKIP key will jump directly to D10. Holding down the HOLD key will stop the scroll at the
current display.
D#
DISPLAY
EXPLANATION OF DISPLAY CONTENTS
D1
LOADING
SENSOR
Indicates the information from the sensor is being loaded into
the memory of the instrument
D2
Self Check
Indicates the unit is performing electronic checks
D3
Self Check
Passed
Indicates startup checks successfully completed
D4
100
PPM
NH3
HOLD NEXT SKIP
Indicates programmed full scale range.
Bottom line indicates switch functions.
D5
C
20
PPM
↓
NH3
HOLD NEXT SKIP
Indicates the Caution setpoint. The minus sign under the C
indicates that the caution is set at -20 PPM and the
↓
indicates the alarm will activate on a falling signal.
D6
W
25
PPM
↑
NH3
HOLD NEXT SKIP
Indicates the Warning setpoint. The current setpoint is 25
PPM and the
↑
indicates that the alarm will activate on a
rising signal.
D7
A
50
PPM
↑
NH3
HOLD NEXT SKIP
Indicates the Alarm setpoint. The current setpoint is 50 PPM
and the
↑
indicates that the alarm will activate on a rising
signal.
D8
Range 50 PPM
to 500 PPM
HOLD NEXT SKIP
Indicates the minimum and maximum full scale range for the
sensor module currently installed in the detector.
D9
Firmware Rev x.xx
Sensor Rev x.xx
HOLD NEXT SKIP
Indicates the revision levels for the firmware in the C16
detector and the sensor module currently installed.
D10
P
0
PPM
NH3
OFF MENU SMPL
Indicates the current gas concentration. This is the normal
operating display. The battery symbol in the upper left corner
indicates which battery is currently providing power.