aspida
– User Manual
Document Ref: PGA-801-10
-
December 2014
Page 25
11) Push the tool and sensor gently downwards into the case until the sensor locates firmly within
its socket.
12) Hold the top ring of the extraction tool gently, without squeezing the grip (Figure 26) and lift
the tool vertically upwards to remove the extraction tool whilst leaving the new sensor in place.
(A gentle twisting of the tool may help to separate the tool from the sensor).
Figure 26 - Remove the extraction tool
13) Replace the sensor cap and fit the 4 cross-head screws (including washers) into the rear of
the case as shown in Figure 27.
Figure 27 - Replace the sensor cap and secure the screws
WARNING:
Any new sensor fitted to a device will take time to settle to a stable reading. For
this reason, once the sensor has been fitted, the device should be left un-powered for at least
2 hours before attempting to power-up and calibrate.
Once the sensor has been allowed to settle, replace the batteries and fit the battery cover. Power the
device and allow the sensor(s) to warm up. The new oxygen sensor will require calibration and may be
showing a fault due to an over-range reading (this is normal for a new sensor). To calibrate the oxygen
sensor, follow the instructions in the section ‘Oxygen fresh air calibration’.
Summary of Contents for aspida
Page 1: ...aspida User Manual...