7) Lift the extraction tool vertically away from the device and the oxygen sensor should lift out of
its socket as shown in Figure 22.
Figure 22 - Removing the oxygen sensor
8) Remove the old sensor from the extraction tool by gently pushing the sensor from the wide,
top end of the extraction tool.
9) Remove the new sensor from its packaging including the protective paper ring on the foam
seal.
Figure 23 - Insert the new sensor into the extraction tool
10) Insert the new sensor as show in Figure 23, ensuring that the blue plastic pin of the new
sensor is aligned with the extraction tool alignment rib, as shown in Figure 24.