Press the
PC Control
button on the keypad to switch to PC Control Mode – the Auto
Mode indicator will light when the KST is in PC Control Mode
Open
AnagoSharp
software
Select “Home” on the menu bar
Click the “Auto Fill” button to automatically fill the test information with the information from
the most recent test using the drop-down lists.
Change any field as required. Use the drop-down lists to select previously entered data.
Pull the bottom end of the test media and clamp it with a 2.00-2.30kg force reading on the
LCD display. The test must begin within this range.
Start the test by clicking the “Run” button, the test will run.
The test can be stopped at any time by clicking the “Stop test” button.
A screen will pop up asking you if you wish to save the test.
Click “Yes” to save the test
.
Unclamp the test media before retracting the knife.