The pump should be in a place where it is easy to connect or disconnect
the power plug.
4
Always disconnect the power adapter and turn off the pump before
cleaning or repairing. After cleaning, make sure the power adapter is
completely dry before plugging into an electrical outlet.
5
Patients with diabetic disorders are subject to enhanced surveillance.
Hypoglycemia, diabetic patients with under or delay feeding may cause
coma.
6
After the feeding set is loaded, the operator needs to confirm that the
feeding set is correctly installed. The liquid in the feeding set should not
flow by itself, otherwise stop using it and contact customer service.
7
Do not use nails or sharp objects to operate the buttons and screen. It
may damage the pump.
8
Store the pump in accordance with the storage conditions in this
instruction for use.
9
The user should set the feeding parameters in accordance with the
requirements of the physician. The wrong feeding parameters may cause
injury to the patient.
10
When the pump generates high priority alarm, it will automatically stop
running. After removing the alarm source, press the "Start/Stop" button to
continue running. Refer to Section 8. Alarm.
11
The pump should be wiped regularly to ensure that it is clean. Blocking
the sensors will cause detection failure or false alarms. Refer to Section
12.1 Maintenance.
12
The pump and accessories are designed to provide a minimum of 5 years
shelf life.
13
Use Amsino specified power adapter. Do not use unknown power adapter.
14
The power adapter cord, feeding set tubing, and pump accessories may
cause a tripping hazard. Avoid leaving wires, cords, or tubing in a pathway
where a person could trip and sustain an injury.
15
Small foam bubbles may accumulate and may create a maximum volume
of 1.6mL without being detected by the pump, which corresponds to a
maximum length of 23cm air in the tubing. Patients who are extremely
sensitive to receiving air in the stomach during enteral feeding should be
provided with enhanced surveillance.
16
This device complies with Part 15 of the FCC. Operation is subject to the
following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
Any changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses