page. Touch the
orange
number to edit KTO time and then touch "Start KTO".
The feeding task stops and KTO runs.
Fig 7-2-8-3 Manual KTO for Continuous Mode
Fig 7-2-8-4 Manual KTO for Intermittent Mode
When KTO time reached, the pump will continue the unfinished feeding task.
Note:
•
For manual KTO, KTO time can be set from 5 ~ 240 mins.
•
For automatic KTO under intermittent mode, KTO time equals to the
interval waiting time.
•
If KTO time/Interval Waiting time>30mins, 0.2ml bolus is run every 30
mins.
•
If KTO time/Interval Waiting time
≤
30mins, 0.2mL bolus is run one time in
the middle of time window. For example, when KTO time is set as 20 mins,
0.2mL bolus is run when KTO time goes 10 mins.