![Amsino AMSure User Manual Download Page 35](http://html1.mh-extra.com/html/amsino/amsure/amsure_user-manual_2936262035.webp)
File
Name
IFU of
AMSure
®
Enteral
Feeding
Pump and Feeding Set
REF
#
L10196_US
Version
B
Page
35
of
81
AMSINO Confidential
For example, as the page above, when the last 1mL is fed, the 3
rd
feeding is
finished, the pump start interval waiting. The page indicate the Feed 3 is
completed from the programed 8 feeds; the next task Feed 4 will start when
time count down to zero.
Caution
KTO function will be performed during interval waiting if KT
O between meals
is switched on in system setting (Refer to 7.2.8 KTO)
7.2.7 Feeding Complete and Switch Off
When the fed volume reaches the dose setting volume, the system triggers
the feeding complete alarm. Press the “Alarm Reset” button to eliminate the
feeding complete alarm. The pump returns to the last setting page and then
the next feeding task can be prepared.
7.2.7-1 Feeding Complete
Notes:
1) Continuous mode, the fed volume reaches dose setting volume, the
feeding complete alarm will occur (Refer to 8.5 Feeding Complete).
2) Intermittent mode, when all the feeds are finished, the feeding complete
alarm will occur (Refer to 8.5 Feeding Complete).
3) Intermittent mode, when each single feeding is finished, the feeding
complete will not occur. However, an
alarm “Single Dose Complete” will
occur and remind user to flush the tubing per protocol
,
which will not
interrupt the programmed intermittent feeding task.
When user touches “Alarm Reset”, system will return to the interval waiting
page. Refer to 8.14 Single Dose Complete Alarm
When the feeding task is completed. Unload the feeding set. Press the